Model 5000A225A
Rev B
19
2.4.1.5.11
Remote Screen
When the Keylock switch is set to the REMOTE position the Remote screen appears. This screen has all the
same information as the Main Menu screen except that the blue bar graph and associated value are only for
the RF Gain value. Because of this, the ALC Threshold, ALC Detector Gain, and ALC Response values are
displayed separately in black toward the center of the screen.
Figure 2-16. Remote Screen
2.4.1.5.12
Fault Screen
The Fault screen will appear anytime that a fault condition is met. The name of the fault is shown toward the
center of the screen. In either LOCAL or REMOTE Keylock switch positions, a Reset button will appear
allowing local resetting of the fault condition. Faults can also be reset remotely. If the fault reset is
successful, the screen will return to the Main Menu or Remote screen depending on the Keylock switch
position.
Figure 2-17. Fault Screen