3. Application Wiring Diagrams
3.1. SDI-12 Digital Interface (always on)
In this configuration, the BlackBox is permanently on and SDI-12 data is instantly available
on request. See section
for details of the SDI-12 data communication protocol.
Six Core Cable Usage
Core Colour
Designation
Connection
Red
Power ve
10V – 14V DC +ve power supply input
Black
Power supply –ve
Power supply ground
Blue
SDI-12
Controller SDI-12 I/O line
Green
RS485 +
Must be isolated and left floating
White
On/Off switch line
+ve power supply input
Yellow
SDI-12 / Modbus
Power supply ground
Screen
Screen / Shield
Power supply or chassis ground
In this configuration, power should be left on the BlackBox continuously.
If the controller has a switched output to control external devices, this can be connected to
the white core (on/off switch line) and can be pulled to ground to put the BlackBox into
sleep mode if power saving is required.
Please note:
the white core must not be left floating
.
If an AP-7000 is connected, see section
8.4.23 Important note concerning AP-7000 self