8.4.6. AP-300 with Depth
Sequence initiated with Start Measurement command:
Command
Response
<a>M!
or
<a>MC!
<a>0009<CR><LF>
<a>D0!
<a><baro><temp><cond><cond20><cond25><CR><LF>
or
<a><baro><temp><cond><cond20><cond25><crc><CR><LF>
<a>D1!
<a><res><sal><tds><ssg><CR><LF>
or
<a><res><sal><tds><ssg><crc><CR><LF>
Sequences initiated with Additional Measurement command:
Command
Response
<a>M1!
or
<a>MC1!
<a>0003<CR><LF>
<a>D0!
<a><do><do-sat><depth><CR><LF>
or
<a><do><do-sat><depth><crc><CR><LF>
Command
Response
<a>M2!
up to
<a>M9!
or
<a>MC2!
up to
<a>MC9!
<a>0000<CR><LF>
Sequence initiated with Start Concurrent Measurement command:
Command
Response
<a>C!
or
<a>CC!
<a>00012<CR><LF>
<a>D0!
<a><baro><temp><cond><cond20><cond25><res><sal><tds><ssg><do>
<do-sat><CR><LF>
or
<a><baro><temp><cond><cond20><cond25><res><sal><tds><ssg><do>
<do-sat><crc><CR><LF>
<a>D1!
<a><depth><CR><LF>
or
<a><depth><crc><CR><LF>
Sequences initiated with Additional Concurrent Measurement command:
Command
Response
<a>C1!
up to
<a>C9!
or
<a>CC1!
up to
<a>CC9!
<a>00000<CR><LF>