![Aquaread BlackBox Instruction Manual Download Page 56](http://html.mh-extra.com/html/aquaread/blackbox/blackbox_instruction-manual_2960649056.webp)
Aquaread BlackBox™ Instruction Manual
10401-00618 Rev P
AP-2000 with Depth and AP-700/800 with Depth (software version 4.00 and above)
continued...
Sequences initiated with Continuous Measurement command:
Command
Response
<a>R0!
or
<a>RC0!
<a><baro><temp><ph><orp><cond><cond20><cond25><res><sal><tds>
<ssg><CR><LF>
or
<a><baro><temp><ph><orp><cond><cond20><cond25><res><sal><tds>
<ssg><crc><CR><LF>
Command
Response
<a>R1!
or
<a>RC1!
<a><do><do-sat><aux1><aux2><nh3><depth><CR><LF>
or
<a><do><do-sat><aux1><aux2><nh3><depth><crc><CR><LF>
Command
Response
<a>R2!
to
<a>R9!
or
<a>RC2!
to
<a>RC9!
<a><CR><LF>
or
<a><crc><CR><LF>
Note: AP-700s running software version 4.00 and above will output in this format with
invalid data in the <aux1>, <aux2> and <nh3> fields.
AP-800s running software version 4.00 and above will output in this format with Turbidity
data in the <aux1> field and invalid data in the <aux2> and <nh3> fields.
© 2019 Aquaread Ltd.
www.aquaread.com
Page 56 of 71