Aquaread BlackBox™ Instruction Manual
10401-00618 Rev P
8.4.12.
AP-600 with Depth
Sequence initiated with Start Measurement command:
Command
Response
<a>M!
or
<a>MC!
<a>0009<CR><LF>
<a>D0!
<a><baro><temp><ph><turb><cond><CR><LF>
or
<a><baro><temp><ph><turb><cond><crc><CR><LF>
<a>D1!
<a><cond20><cond25><res><sal><CR><LF>
or
<a><cond20><cond25><res><sal><crc><CR><LF>
Sequences initiated with Additional Measurement command:
Command
Response
<a>M1!
or
<a>MC1!
<a>0005<CR><LF>
<a>D0!
<a><tds><ssg><do><do-sat><depth><CR><LF>
or
<a><tds><ssg><do><do-sat><depth><crc><CR><LF>
Command
Response
<a>M2!
up to
<a>M9!
or
<a>MC2!
up to
<a>MC9!
<a>0000<CR><LF>
Sequence initiated with Start Concurrent Measurement command:
Command
Response
<a>C!
or
<a>CC!
<a>00014<CR><LF>
<a>D0!
<a><baro><temp><ph><turb><cond><cond20><cond25><res><sal><tds>
<ssg><CR><LF>
or
<a><baro><temp><ph><turb><cond><cond20><cond25><res><sal><tds>
<ssg><crc><CR><LF>
<a>D1!
<a><do><do-sat><depth><CR><LF>
or
<a><do><do-sat><depth><crc><CR><LF>
Sequences initiated with Additional Concurrent Measurement command:
Command
Response
<a>C1!
up to
<a>C9!
or
<a>CC1!
up to
<a>CC9!
<a>00000<CR><LF>
© 2019 Aquaread Ltd.
www.aquaread.com
Page 41 of 71