Aquaread BlackBox™ Instruction Manual
10401-00618 Rev P
AP-2000 and AP-700/800 (software version 4.00 and above) continued...
Sequences initiated with Continuous Measurement command:
Command
Response
<a>R0!
or
<a>RC0!
<a><baro><temp><ph><orp><cond><cond20><cond25><res><sal><tds>
<ssg><CR><LF>
or
<a><baro><temp><ph><orp><cond><cond20><cond25><res><sal><tds>
<ssg><crc><CR><LF>
Command
Response
<a>R1!
or
<a>RC1!
<a><do><do-sat><aux1><aux2><nh3><CR><LF>
or
<a><do><do-sat><aux1><aux2><nh3><crc><CR><LF>
Command
Response
<a>R2!
to
<a>R9!
or
<a>RC2!
to
<a>RC9!
<a><CR><LF>
or
<a><crc><CR><LF>
Note: AP-700s running software version 4.00 and above will output in this format with
invalid data in the <aux1>, <aux2> and <nh3> fields.
AP-800s running software version 4.00 and above will output in this format with Turbidity
data in the <aux1> field and invalid data in the <aux2> and <nh3> fields.