SD 310_2 07/2018
Error messages during calibrations
> [AL <
flashes in the
upper display
The pre-adjusted calibration interval has
lapsed or the last calibration was invalid
The device must be
calibrated
[AL ERR.1
Incorrect reference point in air
Check sensor and gas/
solution
[AL ERR.2
Transconductance is too low
Incorrect test gas / solution
Check sensor and gas/
solution
Defective sensor element
Replace sensor element
[AL ERR.3
Transconductance is too high
Incorrect test gas / solution
Check sensor and gas/
solution
Defective sensor element
Replace sensor element
[AL ERR.4
Incorrect calibration temperature
Calibration is only possible
in the range from 5 - 40°C
[AL ERR.5
Zero value is too low
Defective sensor element
Replace sensor element
[AL ERR.6
Zero value is too high
Incorrect test gas / solution
Check sensor and gas/
solution
Defective sensor element
Replace sensor element
[AL ERR.7
Pressure measurement error
Check pressure
[AL ERR.8
Signal instable/timeout
Check sensor and gas
pressure
[AL ERR.9
Sensor not recognized: cannot be
calibrated
Check sensor and cabling