Menu
Parameter
Values
Meaning
see
SET
(OR R
Adjusts the measurement channels
0FFS
°C or °F
-5.0 °C to 5.0 °C
or
-9.0 °F to 9.0 °F
The zero point of the temperature
measurement is offset by the set
value. The sensor differences and the
measurement device differences can
both be offset
OFF
Zero point offset is deactivated
(=0.0°)
S[AL
°C or °F
-5.00 - 5.00 % The pitch of the temperature
measurement is modified by this
factor (in %) to offset the sensor and
measurement device differences.
OFF
Pitch correction is deactivated (=0.00)
0FFS
hPa
-20 - 20 hPa
The zero point of the absolute pressure
measurement is offset by the set value.
Sensor differences can thus be offset.
OFF
Zero point offset is deactivated (=0.0°)
SET
AL
Set Alarm: Adjustment of the alarm function
AL. 1
ON / NO.SO
Oxygen monitoring: Alarm on with
horn / alarm on without horn
OFF
No alarm monitoring for oxygen
AL.,N
[ONC
Oxygen alarm channel: Concentration
in %
P. 02
Oxygen alarm channel: Partial pressure
in hPa or mmHg
A 1.LO
e . g . 0 . 0 t o
100.0 %
Oxygen min. alarm threshold (not with
AL. 1. oFF)
A 1.H,
e . g . 0 . 0 t o
100.0 %
Oxygen max. alarm threshold (not with
AlL 1. oFF)
AL. 2
ON / NO.SO
Temperature measurement alarm on
with horn / alarm on without horn
OFF
No alarm function for temperature
measurement
A2.LO
-5.0 to +50.0
°C
Temperature min. alarm threshold (not
with Al. 2. oFF)
(*) If there is data in the logger memory, parameters marked with (*) cannot be opened.
If they should be changed, the data must first be deleted!