Elements to identify the Aqualogic 2000™ Status Display Board are:
See Fig. C on previous page.
1. SW-5 Display Primary Time Cycle
2. SW-3 Set UV Alarm
3. SW-2 Reset Time/Cycle Counter
4. SW-8 DIP Switch
5. SW-7 DIP Switch
6. R25 UV Sensor Set Potentiometer
-
not used
7. J5 UV Coax Connector
-
not used
8. J3 Thermistor In
- not used
9. J9 (Flow In) Temperature Signal
10. J10 (Spare In) UV Signal
Detector:
The Detector is a combination temperature and UV probe. A state-of-the-art
GaN photodiode and semiconductor temperature sensor is uniquely
packaged into a hermetically sealed stainless steel housing. The assembly
of the unit is done under dry nitrogen gas to eliminate the presence of
oxygen and water. The semiconductor sensor signals are then processed
within the sensor and sent via the cable into the Monitoring Station. Do not
attempt to remove any of the screws that seal these ports. By doing so, you
may damage the Detector and will void the warranty and the optional NIST-
traceable calibration certification (if included).
See below.
The Detector is
provided with an O-ring. This O-ring is necessary to create the proper seal
when the unit is installed in the Aquafine
®
UV treatment chamber. Be sure
that the O-ring is present when installing the Detector in the UV treatment
chamber.
Cable:
The cable provided with the Detector interconnects the Detector with the
Monitoring Station. On one end it has an IP-65 grade connector that is
connected to the Detector and the other end is wired directly to the Status
Display Board.