MANUAL KEY Switch
Triggers the CSP unit when Manual Key mode is selected. MANUAL KEY switch indicator
will flash briefly on key press.
MAINS POWER Switch
Double-pole switch that also acts as an over-current circuit breaker. It is situated
above the mains input connector. The switch has standard O I positions.
EMERGENCY STOP switch
In the event of an emergency the unit can be shut down. This switch shuts off the
control logic of the charger thus disabling the system. Rotate to reset.
KEY PAD
Navigate menu options
Select / change options
Select an option / sub menu
Exit to previous menu
EOC (End Of Charge) Indicator
The EOC LED is illuminated when the storage capacitors have reached their potential
voltage (~2.5 to 4 kV). In this way the operator can see if the unit is being run faster
than the charger can cope with. If the operator notices loss of data on the recording
system, it may mean the CSP-SNv is firing too quickly, not allowing the capacitors to
charge to their full potential. Slowing the repetition rate down until the EOC LED flashes
will ensure the appropriate voltage is reached across the storage capacitors, also the
correct energy is being discharged in the sound source. The LCD operation screen
displays a histogram of the EOC level for monitoring.
Manual Key Mode must be selected to active manual fire mode.