REAGENT DISK

The API AQUASPIN uses a disk reagent system. The dried reagents 

are packaged in single test amounts in a sealed, polystyrene disk. 

Stainless steel mixing beads in the reaction chambers mix the 

sample water and the dried reagents. Tests for all factors in the 

series are performed at one time. It is not possible to isolate the 

well for a single factor and perform a test for one test factor only. 

Disks should not be 

fi

 lled in the meter chamber.

UNIVERSAL DISK COVER

The black Universal Disk Cover (1719) is placed over the 

disk in the photometer chamber to reduce interference 

from stray light. The disk cover is positioned over the disk by 

aligning the D-shaped hole in the center of the disk over the 

D-shaped hub in the photometer chamber. The disk cover 

should be placed gently on the hub. There is no need to 

fi

 rmly press the disk cover down onto the hub. The test will be 

aborted if the disk cover is not used.

METER CHECK DISK

The Meter Check Disk (AP-1705) serves two purposes:

 

¬

When the Meter Check Disk (MCD) is run as an option from the Disk Series 

menu, the values on the display are compared to the values on the chart 

on the Meter Check Disk package to determine whether the meter is 

reporting results in the expected ranges. 

 

¬

If the Meter Check Disk is inserted in the meter and Rotary Calibration 

is chosen from the Settings Menu, the alignment of the hub and disk is 

evaluated. The results are analyzed and reported as Pass or Fail. 

For use of the Meter Check Disk see page 15.

USB CABLE 

A USB cable connects the API AQUASPIN to a PC. When used with the AC Power 

Adapter, it connects the meter to an AC outlet. 
WARNING: only use the wall adapter that is supplied with the kit. Make no 

substitutions.

BATTERY

A fully charged battery will last for approximately 150 tests under average 

conditions. The battery life will vary based on usage patterns. The meter should 

be turned off after testing to prolong the battery life. The standard life cycle of a 

lithium ion battery is 500 cycles. The battery will fully charge in approximately 10-12 

hours. The battery is designed to be charged overnight and should be charged 

indoors only. The battery is rated at 12 V and 8.1 AH capacity. Power the meter 

from the battery pack or from AC power. The USB cable and AC adapter are used 

to plug the meter into an AC outlet. WARNING: only use the AC adapter supplied 

with this equipment. Do not substitute.

The battery charge status is indicated by the battery icon on the display. The 

battery icon will indicate when the battery charge is full, partial, low, empty or 

charging. The empty battery icon will 

fl

 ash to indicate that meter should be 

connected to AC power source. If the meter continues to be used at low battery 

Summary of Contents for AP-3580

Page 1: ...Simple Precise Digital Water Testing...

Page 2: ...2...

Page 3: ...k Cover 12 Meter Check Disk 12 Usb Cable 12 Battery 12 Settings 13 TROUBLESHOOTING Meter Troubleshooting Guide 14 Meter Check Disk Troubleshooting Guide 15 AQUASPIN Web Application Troubleshooting Gui...

Page 4: ...re register and clicking Create account 1 You will need to enter the serial number of your meter to create your account The serial number can be found on the bottom of your AQUASPIN meter or on your w...

Page 5: ...t time to doing the update and nothing else on your computer to allow the update to not be disrupted PC CONNECTION When the API AQUASPIN meter is connected to a PC via USB cable the onboard touchscree...

Page 6: ...O NOT overfill the disk If the disk is overfilled sample water will flow out of the overflow hole in the center of the disk If this happens be sure to dry the disk completely before running the test O...

Page 7: ...e air more than necessary Do not transport the meter with a disk in the chamber Store disks at 70 80 F 21 27 C METER When a filled disk is placed in the chamber with universal disk cover and the lid i...

Page 8: ...election tap the icon or word on the screen with a fingertip fingernail pencil eraser or stylus Gently wipe smudges from the screen with the Cloth Wipe Do not touch the screen with a sharp object Do n...

Page 9: ...ount enter their email address and tank type under the REGISTER NEW CONSUMER ACCOUNT section 6 Next select the disk type the disk type can be found on the foil packet Then hit RUN TEST button on the c...

Page 10: ...s to turn the meter off NOTE Remove salt residue daily Salt will damage the meter and cause inaccurate results See the CLEANING section in the manual NOTE For water samples over 100 F 38 C subtract 0...

Page 11: ...esh Q104 21 43 23 May Fresh Q104 20 41 22 May Fresh Q104 20 36 22 May Fresh Q104 Test History View Selected Results Delete Selected Results History Settings Select Auto Save Go to Top Go to Bottom Scr...

Page 12: ...Check Disk is inserted in the meter and Rotary Calibration is chosen from the Settings Menu the alignment of the hub and disk is evaluated The results are analyzed and reported as Pass or Fail For us...

Page 13: ...at any time Set Language There is one language option English Tap to exit to the Settings menu Calibration Tap to run an angle calibration to evaluate the alignment of the hub and disk Tap to exit to...

Page 14: ...filled disk or air bubble Solution Fill disk correctly see FILLING Range Error Raw data out of range Contact Support Output Error Decreased light intensity Possibly dirty lens Clean lens see CLEANING...

Page 15: ...heck Disk with water Water is not used during the meter check procedure Range Check Procedure The Meter Check Disk is used to verify the performance of a meter It is not used to recalibrate the meter...

Page 16: ...ion the LED Calibration should be performed before the Angle Calibration Follow the Cleaning procedure on page 21 to clean the light chamber and optic lenses 1 Tap to go to settings 2 Tab Calibration...

Page 17: ...Universal Disk Cover Code 1719 can be used with the API AQUASPIN Keep the chamber clean and dry Gently swab LED and photodiode lenses located around the hub with a cotton swab dampened with streak fr...

Page 18: ...able 3 To express the test result as ppm Ionized Ammonia NH4 subtract the Unionized Ammonia determined in step 2 from the Total Ammonia 10 C 15 C 20 C 25 C pH Fresh water1 Salt water2 Fresh water Salt...

Page 19: ...nm 635 nm Display Color Capacitive Touchscreen 3 5 in 320 x 240 pixel resolution Wavelength Accuracy 2 nm Wavelength Bandwidth 10 typical Photometric Range 2 to 2 AU Photometric Precision 0 01 AU at...

Page 20: ...anada IC Certification FCC ID WAP2001 License IC 7922A 2001 CE Europe Complies with Directive 1999 5 EC MIC Japan 005 101007 KC Korea MSIP CRM Cyp 2001 EMC EU ETSI EN 301489 1 US FCC PART 15 B CAN ICE...

Page 21: ...horough cleaning when necessary Do not use streak free window cleaner or similar cleaners on the touchscreen REPAIRS Should it be necessary to return the meter for repair or servicing pack the meter c...

Page 22: ...ument Failure to do so could result in personal injury or damage to the meter The API AQUASPIN should not be stored or used in a damp or excessively corrosive environment Care should be taken to preve...

Page 23: ...ance LaMotte Company warrants this instrument to be free of defects in parts and workmanship for 2 years from the date of shipment If it should become necessary to return the instrument for service du...

Page 24: ...pifishcare com Mars Fishcare North America Inc 50 E Hamilton St Chalfont PA 18914 Disk US Patent No 8 734 734 AP 3580 MN 07 22 20 Manufactured by LaMotte Company PO Box 329 Chestertown MD 21620 800 34...

Reviews: