S3——distance regulator: the sensor should be placed on the calibration device to extend out the front end. The
geometrical shape and materials of the support should minimize the impact on the distribution of the electrical
field around the front end of the sensor.
VII. Product Performance
The test data of the standard plate electrode is as follows:
The static voltage values under 3 groups of test distances were detected. The standard plate electrode was
600mm*600mm stainless steel electrode. The test distance was the distancebetween the surface of plastic
shell to that of the plate electrode on one side of the detection window of the electrostatic sensor. The light
grey heavy line in the figure is the calibration voltage line when the slope is 1.
-10000
-5000
0
5000
10000
-10000
-5000
0
5000
10000
标 准 电压( 斜 率K=1)
测试距离100mm
测试距离300mm
测试距离500mm
实测电压值(
V)
标 准 平板 电压值( V)
-1000 -800 -600 -400 -200
0
200
400
600
800
1000
-1000
-800
-600
-400
-200
0
200
400
600
800
1000
实测电压值(
V)
标 准 平板 电压值( V)
标 准 电压( 斜 率K=1)
测试距离100mm
测试距离300mm
测试距离500mm
The following two figures are the test data maps under different horizontal test distances when the vertical
test distance is 500mm, standard plate electrode is 600mm*600mm stainless steel electrode and the sensor is
relative to the central position of the detection plate:
Standard voltage (slope K =1)
Test distance 100mm
Test distance 300mm
Test distance 500mm
A
ct
u
al
me
as
u
red
v
o
lt
ag
e
v
al
u
e
Voltage value of standard plate
Standard voltage (slope K =1)
Test distance 100mm
Test distance 300mm
Test distance 500mm
A
ct
u
al
me
as
u
red
v
o
lt
ag
e
v
al
u
e
Voltage value of standard plate