2. On-line mode: 

(Monitoring terminal, integrated power supply and communication software need to be 

purchased additionally) 

For the power output network port, monitoring terminal network port and electrostatic sensor network port 

of 24V power adapter, the network ports of the three devices are both the power port and the communication 

port, and the network ports of the three devices can be commonly used.   

1

 

Connection with monitoring terminal:   

 

IMS Smart Controller

WAN/LAN

P4

P3

P2

P1

POWER

P1-P4 : 1/2-VCC  3-RS485_B  4-RS485_A  5/6-GND  7-C_VTH/HVAL  8-C_ACT/Clean  9-PE(Metal shield)

P1

P2

P4

P3

Red-VTH/HVAL

    

ACT/Clean-Green

ACT

Wrok/Link

NEXT

SCAN

NEXT

LINK

B

SET

S

SCROLL

T

接AC

电源

网线

网线

手持终端

网口适配器

静电传感器

 

 

2

 

Connection with PC:   

SW1

Red-V

TH

/HV

AL

     

ACT/Clean-

Green

SW2

SW4

SW3

SW5

SW6

SW10

SW8

SW9

SW7

SW11

SW13

SW12

SW14

SW15

SW16

P1

P17

P18

P2

P4

P3

P20

P19

P7

P8

P6

P5

P23

P24

P22

P21

P29

P30

P32

P31

P27

P28

P26

P25

P9

P10

P12

P11

P15

P16

P14

P13

V/A Display

IMS Power Supply

PC

一个串口可连接16台机器

静电传感器

静电传感器

静电传感器

通讯转换器

终端电源

网线

网线

 

 

3. Operating steps (refer to the following attached drawings):   

  According  to  the  static  electricity  quantity  and  the  operating  environmentof  the  charged  object,  rotate  the 

circular dial switch with the 3mmdiameter cross screwdriver to select the test distance.   

 According to the static electricity quantity and the operating environmentof the charged object, turn the bar 

dial switch and select the detection gear.   

 According to withstand static voltage value of the protected product, set the safety (alarm) threshold of static 

electricity via the remote controller or monitoring software.   

Notes:  when  the  detected  static  voltage  value  is  within  the  set  threshold  of  static  electricity,  the  threshold 

alarm indicator light is green, and it will be red if the set threshold is exceeded.   

Hand-held terminal 

 

Network cable 

 

Network cable 

 

Electrostatic sensor 

 

Internet access adapter   

 

Connect with AC100-240V 

50/60Hz power supply 

Terminal power supply 

Network cable 

 

Network cable 

 

Electrostatic sensor 

 

Electrostatic sensor 

 

Electrostatic sensor 

 

RJ45 to USB 

communication converter 

 

One serial port can be 
connected with 16 sets of 
machines 

Summary of Contents for AP-YV3302H

Page 1: ...Electrostatic Sensor AP YV3302H Operating Instruction Shanghai Anping Static Technology Co Ltd Floor 3 Building 27 No 69 Guijing Road Shanghai China Tel 021 64517676 Fax 021 64517673 ...

Page 2: ...of Application It can be widely applied in the industries such as electronics photoelectricity and plastics III Product Features 1 With comfortable feel and beautiful appearance 2 Detection distance can be selected via dial switch 3 Wide detection potential range and high detection precision 4 Alarm threshold can be set and red light alarm indicatesexceeding threshold 5 High definition and all dim...

Page 3: ...ar and minimum resolution corresponding to each measuring distance Detection gear Detection distance coding Detection distance Measurement range Resolution Measuring error 1 0 5mm 2000V 1V 5 1 10mm 4000V 2V 2 25mm 10000V 5V 3 50mm 15000V 10V 4 100mm 20000V 10V 5 150mm 20000V 10V 6 200mm 20000V 15V 7 250mm 20000V 20V 2 8 300mm 20000V 10V 9 350mm 20000V 15V A 400mm 20000V 15V B 450mm 20000V 15V 3 C ...

Page 4: ...ay be changed Subject to the real product and please understand that notice cannot be given in advance 4 Technological function of infrared remote control command Blower Bar Sensor CAL SRZ VTH VTH K K IB IB T T FR S BR S FQ FQ IB IB FS FS RES For user SRZ reset zero VTH VTH threshold adjustment with step unit 10V Unlock 1 After unlocking 30seconds delay for active if any button is pressed 2 After ...

Page 5: ...ce of temperature and humidity on electrostatic detection The lower the temperature is the smaller the humidity is the less water is contained in the space and the more easily the surrounding object triboelectric and the greater the interference to the electrostatic detection is 1 The higher the temperature is the higher the humidity is the more water is contained in the space and the more active ...

Page 6: ...connection Temperature and humidity display LCD Monitoring setting of electrostatic sensor Setting of operating time of electrostatic eliminator Monitoring setting of electrostatic eliminator ACI AC indication on monitor green light is ON for long Network connection red light flashes and green light is ON for long Network disconnection red light flashes and green light is OFF 4 line status indicat...

Page 7: ...inal Overall dimension 100 120 25mm L W H Net weight of monitoring terminal about 400g The overall dimension of monitoring terminal is shown as follows 100 120 52 76 87 5 14 14 4 25 IMS Smart Controller WAN LAN P4 P3 P2 P1 POWER P1 P4 1 2 VCC 3 RS485_B 4 RS485_A 5 6 GND 7 C_V TH HVAL 8 C_ACT Clean 9 PE Metal shield P1 P2 P4 P3 Red VTH HVAL ACT Clean Green ACT Wrok Link NEXT SCAN NEXT LINK B SET S ...

Page 8: ...is shown as follows 输出端口 24V直流1 2线输出 三相电源插座 AC100 240V 50 60Hz Mechanical parameters of communication converter Overall dimension 90 28 22mm L W H Net weight of communication converter about 20 5g The overall dimension of communication converter is shown as follows 接口 接口 RJ45 output port 24V DC 1 2 line output Three phase power socket RJ45 port USB port ...

Page 9: ...ng temperature 0 50 Working humidity 30 65 RH VI Product Calibration 1 Calibration environmental conditions The calibration environmental conditions and requirements are as follows a Environmental temperature 20 5 b Relative humidity 30 60 c There is no measurable electrical field magnetic field as well as positive and negative irons around d There is ground wire and the resistance of ground is 10...

Page 10: ...ectrodes should be wrapped with insulating materials the plate area should be large enough and the diameter or side length should be no less than 0 4m Our calibration plate is square stainless steel plate with the dimension of 600mm 600mm R protective resistance the withstand voltage strength of resistance is 20KV and the current through the protective resistance and human body is 5mA and the resi...

Page 11: ...gure is the calibration voltage line when the slope is 1 10000 5000 0 5000 10000 10000 5000 0 5000 10000 标 准 电压 斜 率K 1 测试距离100mm 测试距离300mm 测试距离500mm 实测电压值 V 标 准 平板 电压值 V 1000 800 600 400 200 0 200 400 600 800 1000 1000 800 600 400 200 0 200 400 600 800 1000 实测电压值 V 标 准 平板 电压值 V 标 准 电压 斜 率K 1 测试距离100mm 测试距离300mm 测试距离500mm The following two figures are the test data maps under different horizontal t...

Page 12: ...m 200mm 250mm 300mm 350mm 400mm 450mm 500mm 550mm 600mm 700mm at this point the detection data displayed by the terminal is the most accurate Vertical test distance 500mm Actual measured voltage value Horizontal test distance Baseline Voltage Baseline Voltage Baseline Voltage Baseline Voltage Baseline Voltage Baseline Voltage Baseline Voltage Baseline Voltage Baseline Voltage Baseline Voltage Base...

Page 13: ... PC 一个串口可连接16台机器 静电传感器 静电传感器 静电传感器 转 通讯转换器 终端电源 网线 网线 3 Operating steps refer to the following attached drawings According to the static electricity quantity and the operating environmentof the charged object rotate the circular dial switch with the 3mmdiameter cross screwdriver to select the test distance According to the static electricity quantity and the operating environmentof the charged obj...

Page 14: ...otate the circular dial switch with the 3mmdiameter cross screwdriver to set the equipment address 4 Installation drawing Select test distance Static threshold alarm indicator light Select equipment address Select measurement range for test ...

Page 15: ...4 1 Structure chart of each mounting fixture ...

Page 16: ...re use Before power on please check the specification of the provided power supply Any power supply nonconforming to the specification will cause damage and even failure of product Please operate the product under the specified temperature 0 50 During test the distance between the person and the product should be more than 1m to avoid Φ5 diameter stainless steel mounting stem Round head knurled cr...

Page 17: ...e displayed static value may be zero Don t install high voltage equipment around such as high voltage power supply electrostatic generator iron generator and eliminator because the high voltage will affect the performance and detection accuracy of the sensor During the installation of high vibration area please fasten the sensor otherwise data error may occur Please regularly check the power cable...

Page 18: ...e good performance of the product please store the equipment in a dry place away from light when it s not used Please don t press it heavily 2 This equipment is a precision detector please don t vibrate it intensively during use X After sales Services AP YV3302H electrostatic sensor has received strict testing inspection and aging treatment before delivery and its performance can completely reach ...

Reviews: