Ozone Detector c/w

O3TH

Dec.11

    Automatikprodukter

Indicator for Temperature/Humidity

Table:

Ozone measurement in normal air Y1=0

Sensor output in normal air  X1  

(

zero point)

Ozone measurement in sample gas1 Y2

Sensor output in gas1 X2

Ozone measurement in sample gas2 Y3

Sensor output in gas2 X3

Ozone measurement in sample gas3 Y4

Sensor output in gas3 X4

Ozone measurement in sample gas4 Y5

Sensor output in gas4 X5

Press switch key for about 25 seconds until the monitor is off, then all the settings return to the default.

If warm-up time is less than 3 months before leaving factory, then do emendation as the second row of the following table.

If warm-up time is more than 3 months before leaving factory, then put four DIPs to ON and 

p

ress switch key for about 25

seconds until the detector is off, then do emendation as the third row of the following table.

Warm-up time

0ppb

40ppb

80ppb

200ppb

300ppb

500ppb

600ppb

700ppb

900ppb

999ppb

Less than 3 months

1000

847

717

500

420

338

315

291

264

254

More than 3 months

870

720

580

500

420

338

315

291

264

254

Summary of Contents for O3TH

Page 1: ...Features Real time detection of ambient ozone gas level Full time indicate O3 level Wall mount with LCD display Temperature humidity detection The on off output for control a fan O3 generator or an a...

Page 2: ...t off power first Simualtaneously depress the 2 clips on either of the sides of the detector gently with your nails or other unsharp tools then move off the face part 2 Mount the detector on the place...

Page 3: ...range is 10 1000ppb accuracy 1ppb and setting range is 40 1000ppb its maximum measurement limited is 2000ppb 3 Turn on Turn on the monitor by press power for 2 seconds Temperature measurement displays...

Page 4: ...p on press switch key for about 20 seconds until X10 symbol disappearing It indicates the zero has no any calibration it s still 0 Note Auto zero calibration can be done only when ozone measurement is...

Page 5: ...tion 1 9600bps 2 14400bps 3 3 19200bps 4 28800bps 5 38400bps RESET Put the DIP1 to ON and press switch key for about 25 seconds until the detector is off then all the setting return to the default Ann...

Page 6: ...is finished 6 Now go to demarcation point 2 put the controller in sample gas 7 Write down the un twinkling point of X2 and Y2 8 Press switch to the second demarcation point N 2 9 Press tangent down o...

Page 7: ...key for about 25 seconds until the monitor is off then all the settings return to the default If warm up time is less than 3 months before leaving factory then do emendation as the second row of the...

Page 8: ...ent Controller Placement The Ozone Detector has been designed to measure the ambient concentration of ozone The controller must not be placed directly in an ozone stream For indoor local area monitori...

Reviews: