Ozone Detector c/w
O3TH
Dec.11
Automatikprodukter
Indicator for Temperature/Humidity
Table:
Ozone measurement in normal air Y1=0
Sensor output in normal air X1
(
zero point)
Ozone measurement in sample gas1 Y2
Sensor output in gas1 X2
Ozone measurement in sample gas2 Y3
Sensor output in gas2 X3
Ozone measurement in sample gas3 Y4
Sensor output in gas3 X4
Ozone measurement in sample gas4 Y5
Sensor output in gas4 X5
Press switch key for about 25 seconds until the monitor is off, then all the settings return to the default.
If warm-up time is less than 3 months before leaving factory, then do emendation as the second row of the following table.
If warm-up time is more than 3 months before leaving factory, then put four DIPs to ON and
p
ress switch key for about 25
seconds until the detector is off, then do emendation as the third row of the following table.
Warm-up time
0ppb
40ppb
80ppb
200ppb
300ppb
500ppb
600ppb
700ppb
900ppb
999ppb
Less than 3 months
1000
847
717
500
420
338
315
291
264
254
More than 3 months
870
720
580
500
420
338
315
291
264
254