background image

   Automatikprodukter

Ozone Detector c/w
Indicator for Temperature/Humidity

O3TH

   Dec.11

Steps:

1)  Energize the controller

2)  Put the four DIPs to the upper position to be ON and wait until the digits displayed on the LCD to be stable

3)  Demarcate demarcation point 1 (0), write down the un-twinkling point of X1

4)  Press switch to the first demarcation point (N=1)

5)  Press (tangent down)  or (tangent up) to adjust the twinkling figure until it equals to X1.

Then the demarcation of X1 is finished

6)  Now go to demarcation point 2, put the controller in sample gas

7)  Write down the un-twinkling point of X2 and Y2

8)  Press switch to the second demarcation point (N=2)

9)  Press (tangent down) or to (tangent up) adjust the twinkling figure until it equals to Y2

      10) Press switch, then the following figure starts twinkling

      11) Press (tangent down) or (tangent up) to adjust the twinkling figure until it equals to X2.

     Then the demarcation of X2 is finished

      12) Repeat steps 6-11 to finish the emendation of demarcation point 3 to 10

       13) After the emendation to all demarcation points, put the four DIPs to OFF.

         

  

6

          1000PPB      7

            

5

                                                                     

 4

                                                                             

 

3

                                                                                   

2

                                                                                         

     1   1023              X (A/D transition)

    Y ( ozone measurement)

Summary of Contents for O3TH

Page 1: ...Features Real time detection of ambient ozone gas level Full time indicate O3 level Wall mount with LCD display Temperature humidity detection The on off output for control a fan O3 generator or an a...

Page 2: ...t off power first Simualtaneously depress the 2 clips on either of the sides of the detector gently with your nails or other unsharp tools then move off the face part 2 Mount the detector on the place...

Page 3: ...range is 10 1000ppb accuracy 1ppb and setting range is 40 1000ppb its maximum measurement limited is 2000ppb 3 Turn on Turn on the monitor by press power for 2 seconds Temperature measurement displays...

Page 4: ...p on press switch key for about 20 seconds until X10 symbol disappearing It indicates the zero has no any calibration it s still 0 Note Auto zero calibration can be done only when ozone measurement is...

Page 5: ...tion 1 9600bps 2 14400bps 3 3 19200bps 4 28800bps 5 38400bps RESET Put the DIP1 to ON and press switch key for about 25 seconds until the detector is off then all the setting return to the default Ann...

Page 6: ...is finished 6 Now go to demarcation point 2 put the controller in sample gas 7 Write down the un twinkling point of X2 and Y2 8 Press switch to the second demarcation point N 2 9 Press tangent down o...

Page 7: ...key for about 25 seconds until the monitor is off then all the settings return to the default If warm up time is less than 3 months before leaving factory then do emendation as the second row of the...

Page 8: ...ent Controller Placement The Ozone Detector has been designed to measure the ambient concentration of ozone The controller must not be placed directly in an ozone stream For indoor local area monitori...

Reviews: