12 Installation guide
Some steps of the installation and calibration procedure are described in the
service manual (171.0020) which is provided with this upgrade kit (see
references in this installation guide). The FW upgrade procedure using the FW
wizard in Dialogue is described in the FW upgrade user manual (171.7032).
The standard dummy cell test is described in chapter 13 in the DECADE II
user manual (171.0010). Please make sure that you have the user manual
available during the upgrade procedure.
The complete upgrade procedure will take approximately 2 hours, 30 minutes
to install hardware and 100 minutes to check and recalibrate the DECADE II.
1. Upload of firmware version 3.xx
Upgrade the firmware of the single sensor board DECADE II with an FW version which
supports dual sensor board control (FW > 3.20) as described in the FW upgrade user
guide (171.7032). This step is necessary to be able for the system to recognize the
second board after installation. Note that in step 15 the firmware should be installed for a
second time after the installation of the second sensor board. A CD-ROM is provided with
DECADE II Dialogue and DECADE II firmware release 3.20 >. Please check for new
releases on the internet: http://www.antecleyden.com/support/. If newer versions of either
the FW and Dialogue are available please install these versions.