Page 26
5
1.
Use a coin or the battery cap tool to unscrew the battery cap with a counter-clockwise
motion.
2. Check your battery cap to be sure the colored (yellow) o-ring fits securely and is not torn or
damaged.
3. Check the vent hole on the top of the battery cap to be sure it is clear of debris. This vent
maintains pressurization while preventing water from entering the compartment.
The
battery
cap should be replaced once each year or sooner if the o-ring is damaged or the vent is
clogged. See Chapter 13, Care and Maintenance.
4. Insert the Energizer AA L91 Lithium (1.5V) battery into the battery compartment with the
positive (+) end going in first.
5. Replace the cap by turning clockwise until secure but
do not over tighten
. The cap should be
flush with the pump case and you should not see the colored o-ring.
6. Each time you change the battery, your IR 1250 will run a series of self-tests, which will last
a few seconds. The Animas logo will be displayed briefly. The pump will give a beep to alert
you to
verify
(or change) the time/date, language and battery type.
7. Check the displayed time/date, battery type and language. If correct, scroll down to highlight
Confirm and press the OK button. The Home screen will be displayed. For
more details on changing the time and date, see next section,
Setup - Basics,
Setting/Changing Time and Date.
Note: The time and date must be set to Confirm the Verify screen.
8. To change the battery type, highlight the battery field and press OK to
activate Edit Mode (indicated by flashing cursor).
Energizer AA L91 Lithium battery
Yellow o-ring
Battery Cap Vent Hole
Summary of Contents for IR 1250
Page 1: ...Page 1...
Page 6: ...Page 6...
Page 11: ...Page 11 Radiology Quick Reference Chart 1...
Page 16: ...Page 16...
Page 18: ...Page 18...
Page 20: ...Page 20...
Page 44: ...Page 44...
Page 54: ...Page 54...
Page 62: ...Page 62...
Page 78: ...Page 78...
Page 80: ...Page 80...
Page 92: ...Page 92...
Page 94: ...Page 94...
Page 104: ...Page 104...
Page 113: ...Page 113...