Page 108
Tech Specs
Note: When applicable testing used 23” Comfort™ infusion set and temperature of 23 C +/- 1C
Number of Basal Segments: 12 per Program
Number of Basal Programs: 4
Basal Delivery Frequency (minimum): every 3 minutes
Temp Basal Range: -90% to +200%, in 10% increments, OFF
Temp Basal Duration: 0.5 hrs to 24 hrs
Extended Bolus Duration: 0.5 hrs to 12 hrs; 0.1 hr
Battery Type: 1.5 Volt Energizer AA L91 Lithium
Number of Batteries: 1
Battery Life: approximately 6 to 8 weeks for a lithium battery and approximately 3 to 4 weeks
for an alkaline battery
Maximum volume infused under single fault condition: Max 2.0 U
Cartridge Capacity up to: 2.0 mL or 200 units
Continuous Operation, Internally Powered Device
Type BF Medical Equipment (Patient isolated, not defibrillator protected)
Watertight Equipment, IPX8 (protected against the effects of submersion, tested at 12 feet/ 3.6
meters for 24 hours).
Storage Conditions:
-4ºF (-20ºC) to +140ºF (+60ºC)
10% to 100% relative humidity, including condensing
50 kPa to 106 kPa
Batteries must be removed during storage periods exceeding 2 weeks.
Operating Conditions:
+40ºF (+5ºC) to +104ºF (+40ºC)
Outside these temperatures, the flow accuracy and time to occlusion could be
compromised.
20% to 90% relative humidity, including condensing
Ambient pressure: 70 kPa to 106 kPa
Pump Disposal: Contact Animas Corporation for pump disposal information.
Summary of Contents for IR 1250
Page 1: ...Page 1...
Page 6: ...Page 6...
Page 11: ...Page 11 Radiology Quick Reference Chart 1...
Page 16: ...Page 16...
Page 18: ...Page 18...
Page 20: ...Page 20...
Page 44: ...Page 44...
Page 54: ...Page 54...
Page 62: ...Page 62...
Page 78: ...Page 78...
Page 80: ...Page 80...
Page 92: ...Page 92...
Page 94: ...Page 94...
Page 104: ...Page 104...
Page 113: ...Page 113...