Page 72
12
ezBG
This feature allows you to enter your BG reading and your IR 1250 will
calculate a correction bolus based on the Insulin Sensitivity Factor and BG
Target Range for the current time of day. If the Insulin on Board (IOB)
feature is activated, your IR 1250 will calculate a reduced dose for high
BG correction boluses that you can review.
1. From the
Bolus Menu
, select
ezBG
. Press the OK button.
2. The
Actual
field will be highlighted and flashing to indicate Edit mode.
Use the Up/Down arrows to enter your actual BG reading. Press the OK
button to confirm the entry and exit Edit mode.
3. Make sure the target amount is correct. Your healthcare team will give
you your target BG level. If you need to edit this field, press OK to
activate Edit mode. Use Up/Down arrows to change target. Press OK to
confirm target number and to exit Edit Mode.
4. The Insulin Sensitivity Factor
(ISF)
field is highlighted. Check to be sure it is correct. Your
healthcare team will give you your ISF. If you need to edit this field, press OK to activate Edit
mode. Use Up/Down arrows to change target. Press OK to confirm ISF and to exit Edit mode.
5.
Show Result
is highlighted. Press OK.
Summary of Contents for IR 1250
Page 1: ...Page 1...
Page 6: ...Page 6...
Page 11: ...Page 11 Radiology Quick Reference Chart 1...
Page 16: ...Page 16...
Page 18: ...Page 18...
Page 20: ...Page 20...
Page 44: ...Page 44...
Page 54: ...Page 54...
Page 62: ...Page 62...
Page 78: ...Page 78...
Page 80: ...Page 80...
Page 92: ...Page 92...
Page 94: ...Page 94...
Page 104: ...Page 104...
Page 113: ...Page 113...