Alerts, Warnings and Alarms
14
Warning: Exceeds Max Basal
F
Cause
Basal delivery (or Temp Basal) exceeds user-set
maximum.
Effect
Basal delivery stops.
Message
Displayed when awakened (by basal delivery attempt
every 3 min. or manually) until confirmed.
Action
Press OK to confirm. Reprogram Max Basal amount in
Setup Advanced (or reprogram Temp Basal).
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: Exceeds Insulin Remaining
Cause
Basal or Bolus delivery exceeds insulin remaining in
cartridge.
Effect
Basal or Bolus delivery stopped.
Message
Once per occurrence and each time awakened until
confirmed.
Action
Press OK to confirm. Replace with full cartridge.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Warning: No Prime, No Delivery
F
Cause
Pump is not primed.
Effect
All deliveries stop.
Message
Every 3 minutes or when awakened manually.
Action
Press OK to confirm. Disconnect, reprime.
Beeps/Vib
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vibe within one hour.
Alarm: Occlusion
F
Cause Occlusion
detected.
Effect
All deliveries stop.
Message Continuous
until
confirmed.
Action
Press OK to confirm. Disconnect and prime to clear
occlusion. Option to select Suspend (see Suspend).
Beeps/Vib
User-selected every 3 minutes. If not confirmed,
progresses to sweep/vibe within one hour. (Once
confirmed, No Prime alarm triggered, see No Prime.)
Summary of Contents for IR 1250
Page 1: ...Page 1...
Page 6: ...Page 6...
Page 11: ...Page 11 Radiology Quick Reference Chart 1...
Page 16: ...Page 16...
Page 18: ...Page 18...
Page 20: ...Page 20...
Page 44: ...Page 44...
Page 54: ...Page 54...
Page 62: ...Page 62...
Page 78: ...Page 78...
Page 80: ...Page 80...
Page 92: ...Page 92...
Page 94: ...Page 94...
Page 104: ...Page 104...
Page 113: ...Page 113...