Chapter 13 Care and Maintenance
13
The Vents
Your IR 1250 features a redundant Vent Safety System. Vents serve two purposes. First, they
allow air to pass through the pump so that pressure is equalized under a variety of environmental
circumstances, such as changes in altitude. Second, the vents are backed by a special membrane,
which keeps water from entering the pump.
Battery Cap with O-ring and Vent
Your battery cap contains a yellow o-ring and vent. There is a tiny hole backed by a membrane,
which allows air to pass through but prevent water from entering. The o-ring helps to keep your
pump waterproof. It is recommended that you change the battery cap/vent once a year. If you
work in a dusty environment such as a construction site, mill, cement factory, etc., or if you are a
frequent swimmer, you should change your battery cap every 6 months. Contact Animas
Corporation to order a replacement battery cap.
Under no circumstances should you introduce any kind of sharp object into the vent
openings to clean them. Doing so could compromise the pump’s waterproof capabilities. If at
any time you suspect the vent opening is clogged, replace the battery cap.
Cleaning
Do not use household or industrial cleaners, chemicals, solvents, bleach, scouring pads or
sharp instruments to clean the pump. Never put your pump in the dishwasher or use very hot
water to clean it. Use only a very mild detergent (for example a drop of liquid soap in a glass of
water) and a lint-free cloth.
Never put your pump in a microwave oven or baking oven to dry it. Use a soft towel.
General Wear and Tear
If you drop the pump or it has been hit against something hard, inspect it to be sure it is still
working properly. Check that the display screen is working and clear, that the cartridge cap,
battery cap and infusion set are properly in place. Check for leaks around the cartridge by
wrapping a piece of tissue around the connection area. Cracks, chips or damage to the pump
may impact the battery contact and/or the waterproof feature of your pump. Call our 24/7 Pump
Support Department at 877-767-7373 if you identify or suspect your pump has been damaged.
They will help determine if your pump should be replaced. (Outside the US and Canada, contact
your local Animas distributor.)
Disposal
International regulations require controlled disposal of devices such as insulin pumps. Please
contact your local Animas distributor to ensure proper disposal.
Vent Hole
O-ring
Summary of Contents for IR 1250
Page 1: ...Page 1...
Page 6: ...Page 6...
Page 11: ...Page 11 Radiology Quick Reference Chart 1...
Page 16: ...Page 16...
Page 18: ...Page 18...
Page 20: ...Page 20...
Page 44: ...Page 44...
Page 54: ...Page 54...
Page 62: ...Page 62...
Page 78: ...Page 78...
Page 80: ...Page 80...
Page 92: ...Page 92...
Page 94: ...Page 94...
Page 104: ...Page 104...
Page 113: ...Page 113...