10. DELETING MEASUREMENTS FROM THE MEMORY
When any result is displaying, keep on pressing button ―MEM‖ for three seconds, all results in the memory
bank
will be deleted after three ―beep‖.
See picture 10 & picture 10-1.
Press the button ―MEM‖ or ―START‖,
the monitor will turn off.
Picture 10 Picture 10-1
11. ASSESSING HIGH BLOOD PRESSURE FOR ADULTS
The following guideline for assessing high blood pressure (without regard to age or gender) has been
established by the World Health Organization (WHO). Please note that other factors (e.g. diabetes, obesity,
smoking, etc.) need to be taken into consideration. Consult with your physician for accurate assessment, and
never change your treatment by yourself.
Note: It is not intended to provide a basis of any type of rush toward emergency
conditions/diagnosis based on the color scheme and that the color scheme is meant only to
discriminate between the different levels of blood pressure.
12. TECHNICAL ALARM DESCRIPTION
The monitor will show 'SYS HI','SYS Lo' ,'DIA HI' or 'DIA Lo' as technical alarm on LCD with no delay if the
determined blood pressure (systolic or diastolic) is outside the rated range specified in part SPECIFICACIONS.
In this case, you should consult a physician or check if your operation violated the instructions.
The technical alarm condition (outside the rated range) is preset in the factory and cannot be adjusted or
inactivated. This alarm condition is assigned as low priority according to IEC 60601-1-8.
The technical alarm is non-latching and need no reset. The signal displayed on LCD will disappear automatically
after about 8 seconds.