13.TROUBLESHOOTING (1)
PROBLEM
POSSIBLE CAUSE
SOLUTION
LCD Display shows
battery symbol
Low Battery
Change the batteries
LCD Display shows
―EE‖
Arm or blood pressure monitor
was moved during testing
Re-test taking care to not
move your arm or the blood
pressure monitor
The cuff does not inflate
properly or pressure falls
quickly during testing
Make certain the plugs and
brackets of the cuff are
securely attached to the
monitor.
The cuff was not properly
applied.
Apply the cuff correctly and
try again
No response when you
press button or load
battery.
Incorrect operation or strong
electromagnetic interference.
Take out batteries for five
minutes, and then reinstall all
batteries.
In Clock Mode, no
response when you
press button.
The touch switch is in OFF
position.
Please keep touch switch in
ON position, and try again
14. TROUBLESHOOTING (2)
PROBLEM
POSSIBLE CAUSE
SOLUTION
LCD Display shows
abnormal result
The cuff position was not
correct or it was not properly
tightened
Apply the cuff correctly and try
again
Body posture was not correct
during testing
Review the ―BODY POSTURE
DURING
MEASUREMENT‖
sections of the instructions and
re-test.
Speaking,
arm
or
body
movement, angry, excited or
nervous during testing
Re-test when calm and without
speaking or moving during the
test
Irregular heartbeat
(arrhythmia)
It is inappropriate for people
with serious arrhythmia to use
this Electronic
Sphygmomanometer.