INTENDED USE 

Fully  Automatic  Electronic  Blood  Pressure  Monitor  is  for  use  by  medical  professionals  or  at  home  and  is  a 

non-invasive  blood  pressure  measurement  system  intended  to  measure  the  diastolic  and  systolic  blood 

pressures and pulse rate of an adult  individual by  using a non-invasive technique in  which an  inflatable cuff is 

wrapped around the wrist. 

CONTRAINDICATION 

It is inappropriate for people with serious arrhythmia to use this Electronic Sphygmomanometer. 

PRODUCT DESCRIPTION 

Based on Oscillometric methodology and silicon integrated pressure sensor, blood pressure and pulse rate can 

be measured automatically and non-invasively. The most recent 60 measurements can be stored in the memory 

with  date  and  time  stamp.  The  voice  function  will  ease  the  operation.  The  Electronic  Sphygmomanometer 

corresponds to the below standards:IEC 60601-1:2005/EN 60601-1:2006/AC:2010 (Medical electrical equipment 

--  Part  1:  General  requirements  for  basic  safety  and  essential  performance),  IEC60601-1-2:2007/EN 

60601-1-2:2007 /AC:2010 (Medical electrical equipment -- Part 1-2: General requirements for basic safety and 

essential  performance  -  Collateral  standard:  Electromagnetic  compatibility  -  Requirements  and  tests),  IEC 

80601-2-30

2009+Cor.2010/EN  80601-2-30:2010(Medical  electrical  equipment 

–Part  2-30:  Particular 

requirements 

for 

the 

basic 

safety 

and 

essential 

performance 

of 

automated 

non-invasive 

sphygmomanometers)EN  1060-1:  1995  +  A1:  2002  +  A2:  2009  (Non-invasive  sphygmomanometers  -  Part  1: 

General requirements), EN 1060-3: 1997 + A1: 2005 + A2: 2009 (Non-invasive sphygmomanometers - Part 3: 

Supplementary  requirements  for  electro-mechanical  blood  pressure  measuring  systems),  ANSI/AAMI 

SP-10:2002+A1:2003+A2:2006. 

SPECIFICATIONS 

1.  Product name: Blood Pressure Monitor 

2.  Model: KD-7965 

3.  Classification: Internally powered, Type BF applied part,IPX0,No AP or APG, Continuous operation 

4.  Machine size: 88mm x 66mm x 28.7mm 

5.  Cuff circumference: 14-19.5cm/5.5-7.7inches 

6.  Weight: about 142g (exclude batteries) 

7.  Measuring method: oscillometric method, automatic air inflation and measurement 

8.  Memory volume: 60 times with time and date stamp 

9.  Power source: batteries :2 ×1.5V

SIZE AAA 

10.    Measurement range: 

Cuff pressure: 0-300mmHg 

Summary of Contents for KD-7965

Page 1: ...ERY LOADING 5 2 TOUCH SCREEN SWITCH 6 3 CLOCK AND DATE ADJUSTMENT 6 4 VOICE SETTING 6 5 CONNECTING THE CUFF TO THE MONITOR 7 6 APPLYING THE CUFF 7 7 BODY POSTURE DURING MEASUREMENT 7 8 TAKING YOUR BLO...

Page 2: ...level in the daytime while most people are awake and active Considering the above information it is recommended that you measure your blood pressure at approximately the same time each day Too freque...

Page 3: ...ssential performance IEC60601 1 2 2007 EN 60601 1 2 2007 AC 2010 Medical electrical equipment Part 1 2 General requirements for basic safety and essential performance Collateral standard Electromagnet...

Page 4: ...ressure exceed 300 mmHg or maintained above15 mmHg for longer than 3 minutes of the bladder may cause ecchymoma of your arm 7 Consult your physician if you have any doubt about below cases 1 The appli...

Page 5: ...s B digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates u...

Page 6: ...ly touch screen switch in ON position the month will blink at first See picture 3 1 Press the button START repeatedly the day hour and minute will blink in turn See picture 3 2 3 3 3 4 While the numbe...

Page 7: ...ffed wrist in front of your body on a desk or table with the palm up If the cuff is correctly placed you can read the LCD display c The cuff must be neither too tight nor too loose Note Please refer t...

Page 8: ...y stored in the memory If the voice function is on it will announce the measurement result f After measurement the monitor will turn off automatically after 1 minute of no operation Alternatively you...

Page 9: ...r treatment by yourself Note It is not intended to provide a basis of any type of rush toward emergency conditions diagnosis based on the color scheme and that the color scheme is meant only to discri...

Page 10: ...on or strong electromagnetic interference Take out batteries for five minutes and then reinstall all batteries In Clock Mode no response when you press button The touch switch is in OFF position Pleas...

Page 11: ...iately qualified technical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a mini...

Page 12: ...omagnetic emissions The KD 7965 is intended for use in the electromagnetic environment specified below The customer or the user of the KD 7965 should assure that it is used in such an environment Emis...

Page 13: ...erial the relative humidity should be at least 30 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical loca...

Page 14: ...l situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular...

Page 15: ...r of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7...