Page 6-3
6.5 KEY-IN ZERO CALIBRATION VALUE (F19)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss.
A valid zero calibration value, obtained from a successful F16 calibration procedure, must
be used.
1. While in the Setup mode, scroll to "
F 19
", then scroll down once using the ZERO key.
2. The display will momentarily show "
CAL 0
", followed by a flashing zero. Use the four
directional keys (shown in Figure 6-1) to adjust the displayed value to the zero calibration
value.
3. After setting the exact value, press the NET/GROSS key to save the value.
4. The display will show "
E CAL 0
" momentarily, then revert back up to F19.
6.6 KEY-IN SPAN CALIBRATION VALUE (F20)
Note:
This procedure is intended for emergency use only in the case of non-volatile memory loss.
A valid span calibration value, obtained from a successful F17 calibration procedure, must
be used.
1. While in the Setup mode, scroll to "
F 20
", then scroll down once using the ZERO/OFF key. The
indicator will prompt you to enter the information in Table 6-2.
2. If the value shown is correct, press the ZERO/OFF key to move to the next parameter.
Otherwise, Use the four directional keys (shown in Figure 6-1) to adjust the displayed value to
the span calibration value.
3. After setting the exact value, press the NET/GROSS key to save the value.
4. If the entered values are entered successfully, the display will show "
E
" momentarily before
continuing to the next parameter. At the completion of the sequence, the indicator will then
revert back up to F20.
CODE
NAME
ET T 1
First Test Weight Value
ET C 1
First Span Calibration Value
ET T 2
Second Test Weight Value
ET C 2
Second Span Calibration Value
ET T 3
Third Test Weight Value
ET C 3
Third Span Calibration Value
Table 6-2: Calibration Value Entry Table