Page 5-1 

 

CHAPTER 5: USER MENU DESCRIPTIONS AND PROCEDURES 
 

NOTE:  The UO-200 only supports the A5 (Units Key) and A10 (Auto Power Off) functions. Please 
disregard the other listed functions. 

 

5.1  USER MENU DESCRIPTIONS 

 

This section provides more detailed descriptions of the selections found in the User Menu Chart. 
Factory-set defaults are shown in bold with a checkmark (

). 

 

NAME/CODE DESCRIPTION  CODE/VALUE 

A5

 

Disable the 
UNITS Key

 

Allows the UNITS key to be disabled so that an operator cannot 
accidentally press the key and change the displayed units. 

"0" = Disable the Units key              "1" = Enable the Units key 

1

 

A10

 

Auto Power Off 

 

Actuates the function that allows entry of the desired automatic turn 
off time in minutes. Pressing the 

ZERO

 key to scroll down one level 

begins the sequence.  

“0” = Disabled (Always ON) 

0 - 30 

2

 

 

5.2 

USER MENU PROCEDURES

 

 

 

USER MODE KEY FUNCTIONS

ZERO

GROSS

NET

TARE

PRINT

SET

UNITS

 

 

Figure 5-1: User Menu Key Assignments 

 

5.2.1  Auto Power Off Entry (A10) 

 

1.  While in the User Menu mode, scroll to "

A 10

", then scroll down once using the 

ZERO/OFF key to enter the Auto Power Off menu. 

2. 

The display will momentarily show the current line feeds value. 

3.  Use the four directional keys (shown in Figure 5-1 above) to adjust the displayed 

value to the actual auto power off value. Increase the flashing digit by pressing the 
UNITS key. Decrease the flashing digit by pressing the ZERO/OFF key. Pressing 
the ON/PRINT key or the TARE key will change the position of the flashing digit. 

4.  After setting the exact value, press the NET/GROSS key to save the auto power off 

value. The display will show "

SET

" momentarily, then revert back up to A10.  

 

 

Visit www.balances.com Your Authorized A&D Dealer

Summary of Contents for U0-200

Page 1: ...5 2005 A D Weighing Contents subject to change without notice A D Weighing 1555 McCandless Drive Milpitas CA 95131 Tel 800 726 3364 Fax 408 263 0119 E mail scales andweighing com Web www andweighing c...

Page 2: ...in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions manual may cause harmful interference...

Page 3: ...otes on the Setup Menu 3 2 3 2 4 Exiting the Setup 3 2 3 3 User A Menu 3 3 3 3 1 Entering the User Menu 3 3 3 3 2 Navigating in the User Menu 3 3 3 3 3 Notes on the User Menu 3 4 3 3 4 Exiting the Use...

Page 4: ...on Value F19 6 3 6 6 Key in Span Calibration Value F20 6 3 Chapter 7 Operation 7 1 7 1 Display 7 1 7 1 1 Liquid Crystal Display LCD 7 1 7 2 Keyboard 7 2 7 2 1 Function Keys 7 2 7 3 General Scale Opera...

Page 5: ...ntered via the front panel keys including calibration The UO 200 is powered with six C cell alkaline batteries not included An AC adapter option is also available Prior to using the UO 200 please read...

Page 6: ...ector into the load cell port on the rear panel of the indicator 2 1 2 CONNECTING THE POWER SUPPLY 1 Obtain six 6 alkaline C UM 2 batteries and install them into the battery compartment located at the...

Page 7: ...opposite position See Chapter 2 for location of the Setup Calibration Switch 3 Power on the UO 200 The indicator shows F 1 to indicate that you are in Setup Menu mode 3 2 3 NAVIGATING IN THE SETUP MEN...

Page 8: ...o begin Figure 3 2 Setup Menu Chart 3 2 4 NOTES ON THE SETUP MENU 1 There is also an F21 sub menu present that is for FACTORY USE ONLY 2 Detailed descriptions of the setup menu parameters can be found...

Page 9: ...Chart shown in Figure 3 4 on the following page 1 To move to a new A heading use the TARE left or ON PRINT right key to move right or left in the User Menu Chart 2 To move to the selection level pres...

Page 10: ...Enable 0 1 A16 Time Format 0 1 A17 Time Entry A18 Data Enable 0 1 A19 Date Format 0 1 A20 Date Entry Press ZERO key to begin Press ZERO key to begin 1 2 Figure 3 4 User Menu Chart 3 3 3 NOTES ON THE U...

Page 11: ...evel at which motion is detected by comparing the present display update with the previous one If motion is not detected for two seconds or more scale is in standstill and can process a Print or Zero...

Page 12: ...1 Factory Reset This sub menu will reset all parameters in the F and A menu to the default settings USE WITH CAUTION Press the ZERO key twice to exe cute 4 2 SETUP MENU PROCEDURES This section provide...

Page 13: ...down one level begins the sequence 0 Disabled Always ON 0 30 2 5 2 USER MENU PROCEDURES USER MODE KEY FUNCTIONS ZERO GROSS NET TARE PRINT SET UNITS Figure 5 1 User Menu Key Assignments 5 2 1 Auto Powe...

Page 14: ...oll down once using the ZERO key to enter zero calibration menu The display will momentarily show C 0 followed by a value This value is the internal A D count and can prove useful when trying to troub...

Page 15: ...in weight is smaller than 1 of the full capacity of the scale Change the calibration test weight or check the input data Err2 The internal resolution of the scale is not high enough to accept the cal...

Page 16: ...e obtained from a successful F17 calibration procedure must be used 1 While in the Setup mode scroll to F 20 then scroll down once using the ZERO OFF key The indicator will prompt you to enter the inf...

Page 17: ...t is active when ever the displayed weight is within 0 25 divisions of true zero N Indicates that the indicator is displaying net weight G Indicates that the indicator is displaying gross weight T Ind...

Page 18: ...ndicator between Gross weight and Net weight only if a Tare has been established Tare This key is used to establish a Tare provided the following conditions are met 1 The indicator is not at or below...

Page 19: ...y press the ZERO OFF key to obtain a weight reading of zero 3 Place the empty container on the scale s platter and allow the weight indication to stabi lize 4 Press the TARE key The display shows zero...

Page 20: ...rage in EEPROM Excitation Voltage 3 3 VDC 4 x 350 load cells DIGITAL SPECIFICATIONS Microcontroller TI MSP430F447 Program Memory 32K x 8 internal to C SRAM 1Kx 8 internal to C EEPROM 256 x 8 external...

Page 21: ...ration Mode F17 Keyed in weight value is less than 1 of full scale capacity Use a larger test weight or check keyed in value Err 2 Span Calibration Mode F17 There is not enough load cell signal to pro...

Reviews: