Page 4-2
F16
Zero Calibra-
tion
Places indicator into the zero calibration routine. Scrolling down with
the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F17
Span Calibra-
tion
Places indicator into the span calibration routine. Scrolling down with
the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F18
View Calibra-
tion
Actuates the function that allows you to view both the zero and span
calibration value. The values displayed in this function are valid only
after Calibration (F16 & F17) has been successfully completed.
Scrolling down with the
ZERO
key one level begins the procedure.
Press
ZERO
key
to begin sequence
F19
Key-in Zero
Allows you to key-in known zero calibration value in case of memory
loss in the field. Scrolling down with the
ZERO
key one level begins
the procedure.
Press
ZERO
key
to begin sequence
F20
Key-in Span
Allows you to key-in a known span calibration value in case of mem-
ory loss in the field. Scrolling down with the
ZERO
key one level be-
gins the procedure.
Press
ZERO
key
to begin sequence
F21
Factory Reset
This sub-menu will reset all parameters in the “F” and “A” menu to the
default settings. USE WITH CAUTION!
Press the
ZERO
key twice to exe-
cute.
4.2 SETUP MENU PROCEDURES
This section provides instructions for all of the Setup Menu procedures except for the calibration
related menus, which are documented in Chapter 6.
4.2.1 Graduation Entry (F1)
1. While in the Setup Menu mode, scroll to "
F 1
", then scroll down once using the
ZERO/OFF key to enter the Graduation menu.
2. The display will display a value with one flashing digit. This value will be the current
graduation value.
3. Use the four directional keys (shown in Figure 4-1 below) to adjust the displayed
value to the actual graduation value. Increase the flashing digit by pressing the
UNITS key. Decrease the flashing digit by pressing the ZERO/OFF key. Pressing
the ON/PRINT key or the TARE key will change the position of the flashing digit.
SETUP MODE KEY FUNCTIONS
ZERO
GROSS
NET
TARE
SET
UNITS
Figure 4-1: Setup Menu Key Assignments
4.
After setting the exact value, press the NET/GROSS key to save the graduation
value. The display will show "
SET
" momentarily, then revert back up to F1.
NOTE
: The indicator will accept values only in the range from 100 to 50000.