AD-4430B
- 45 -
7.6.
The Parameter List For The Function List
When performing maintenance, use the following list as a memorandum.
When making inquiries about the product, inform your local A&D dealer of the user settings.
7.6.1.
The Calibration Function
(
)
Function No.
Parameter
and range
Item name
Description
Default
value
User
setting MEMO
0 to 0.0000
Decimal point
position
Decimal point position of the weight value
0 0.0 0.00 0.000 0.0000
0
1 to 50
Minimum
division (A
scale / digit )
Minimum division (d) of the weight value
1
2
5
10
20
50
1
1 to
99999
Weighing
capacity
Weighing capacity of the module. Weighing is
possible up to the value of this setting plus 8 d. If
the value exceeds this, overflow will occur and
will not be displayed. The decimal point position
is the same as the setting of
.
70000
0 to 100 Zero range
Range to enable zero adjustment by the
(ZERO)
key Expressed as a percentage of
the weighing capacity with the calibration zero
point as the center. For example, if this is set to 2,
the value in the range of ±2% of the weighing
capacity with the calibration zero point at the
center will be to zero. When a power-ON zero is
performed, the initial zero point will be the center.
2
0.0 to 5.0
Zero tracking
time
Performs zero tracking using this setting in
combination with the setting of the zero tracking
width
. When 0.0, zero tracking will not
be performed. Unit
:
Second in 0.1 increments.
0.0
0.0 to 9.9
Zero tracking
width
Performs zero tracking using this setting in
combination with the setting of the zero tracking
time
. When 0.0, zero tracking will not be
performed. Unit
:
d (minimum division) in 0.1
increments.
0.0
0.0 to 9.9
Stability
detection time
Performs stability detection using this setting in
combination with the setting of the stability
detection width
. When 0.0, stability
detection will not be performed. (Stable all the
time) Unit
:
Second in 0.1 increments.
1.0
0 to 9
Stability
detection
width
Performs stability detection using this setting in
combination with the setting of the stability
detection time
. When 0, stability
detection will not be performed. (Stable all the
time) Unit
:
d (minimum division).
2
0 to 1
Tare and zero
adjustment
when unstable
Tare and zero adjustment when the weight
value is unstable.
0
:
Disables both functions.
1
:
Enables both functions.
1
0 to 1
Tare when the
gross weight
is negative
Tare when the gross weight is negative.
0
:
Disables tare.
1
:
Enables tare.
1