AD-4430B
- 29 -
5.3.7.
The Linearization Function
(
)
Confirm and change linearity settings.
To use this function, select
in the same way as calibration-related functions are
selected.
Function No.
Parameter
and range
Item name
Description
Default
value
0 to 5
Number of input
points
Number of points where linear input was done.
The linear-zero input is included as one point.
Digital linearization is not performed when the set
value is between 1 and 2.
0
-7.0000
to 7.0000
Linear-zero
Voltage for linear-zero input. Unit
:
mV/V.
0.0000
0 to 99999
Linear 1
Mass value
The mass value for linear 1 input.
The decimal point position depends on
.
0
0.0000
to 9.9999
Linear 1
Span voltage
The span voltage between linear-zero and linear 1
input. Unit
:
mV/V.
0.0000
0 to 99999
Linear 2
Mass value
The mass value for linear 2 input.
The decimal point position depends on
.
0
0.0000
to 9.9999
Linear 2
Span voltage
The span voltage between linear-zero and linear 2
input. Unit
:
mV/V.
0.0000
0 to 99999
Linear 3
Mass value
The mass value for linear 3 input.
The decimal point position depends on
.
0
0.0000
to 9.9999
Linear 3
Span voltage
The span voltage between linear-zero and linear 3
input. Unit
:
mV/V.
0.0000
0 to 99999
Linear 4
Mass value
The mass value for linear 4 input.
The decimal point position depends on
.
0
0.0000 to
9.9999
Linear 4
Span voltage
The span voltage between linear-zero and linear 4
input. Unit
:
mV/V.
0.0000