- 34 -
AD-4430B
Function No.
Parameter
and range
Item name
Description
Default
value
1 to 2
Comparison
mass at near-
zero
Item to be compared with near-zero.
1
:
Gross weight
2
:
Net weight
1
-99999 to 99999
Upper limit
value
Reference value for the upper limit.
Decimal point position is linked to
.
10
-99999 to 99999
Lower limit
value
Reference value for the lower limit.
Decimal point position is linked to
.
-10
1 to 2
Comparison
mass of upper
and lower limit
Item to be compared with the upper and lower limit.
1
:
Gross weight
2
:
Net weight
1
1 to 2
Output logic of
upper and
lower limit
Logic used when the result of the comparison with
the upper and lower limit is output
1
:
Positive logic
2
:
Negative logic
1
5.4.4.
The Hold Function
(
)
Function No.
Parameter
and range
Item name
Description
Default
value
0.00 to 9.99
Average time Time to perform the averaging. Unit
:
second.
0.00 is not averaged.
0.00
0.00 to 9.99
Start wait time Time to wait before commencing hold or averaging.
Unit
:
second.
0.00
0 to 2
Condition of
automatic start
Condition to start the hold or averaging automatically.
0
:
Do not use automatic start.
1
:
Above the near-zero range, and stable.
2
:
Above the near-zero range.
0
0 to 1
Release using
control input
Release when control input is falling.
0
:
Do not release
1
:
Release
When the
is set to 0.
Holding
Inputting
Holding
status
When the
is set to 1.
Holding
Inputting
Holding
status
1
0.00 to 9.99
Release time Release after a set amount of time has passed.
Unit : second. 0.00 is not released.
0.00
0 to 99999
Release using
fluctuation range
Release when fluctuation from the holding value
exceeds a set value. The decimal point position is the
same as the setting of
. 0 is not released
.
0
During Holding
During
Holding
During
Holding