![AND AD-4430B Instruction Manual Download Page 29](http://html1.mh-extra.com/html/and/ad-4430b/ad-4430b_instruction-manual_2941312029.webp)
AD-4430B
- 27 -
Function No.
Parameter
and range
Item name
Description
Default
value
0.0 to 9.9
Stability detection
time
Performs stability detection using this setting in
combination with the setting of the stability detection
band
. When 0.0, stability detection will not
be performed. (Stable all the time) Unit
:
Second in
0.1 increments.
1.0
0 to 9
Stability detection
width
Performs stability detection using this setting in
combination with the setting of the stability detection
time
. When 0, stability detection will not be
performed. (Stable all the time) Unit
:
d (minimum
division).
2
Stability detection outputs the STABLE
signal when changes in the weight
value are within a certain range during
a certain time.
0 to 1
Tare and zero
adjustment when
unstable
Tare and zero adjustment when the weight value is
unstable.
0
:
Disables both functions.
1
:
Enables both functions.
1
0 to 1
Tare when the gross
weight is negative
Tare when the gross weight is negative.
0
:
Disables tare.
1
:
Enables tare.
1
0 to 1
Output when
overflow and
unstable
Standard serial output when the weight value
overflows and is unstable.
0
:
Disables output.
1
:
Enables output.
1
1 to 3
Exceeding negative
gross weight
To judge when the negative gross weight is exceeded.
1
:
Gross weight < -99999
2
:
Gross weight < Negative weighing capacity
3
:
Gross weight < -19d
1
1 to 2
Exceeding negative
net weight
To judge when the negative net weight is exceeded.
1
:
Net weight < -99999
2
:
Net weight < Negative weighing capacity
1
0 to 1
Clear the zero value
Select whether or not to clear the zero value.
0
:
Disables.
1
:
Enables.
1
Weight value
Time