![AND AD-4430B Instruction Manual Download Page 19](http://html1.mh-extra.com/html/and/ad-4430b/ad-4430b_instruction-manual_2941312019.webp)
AD-4430B
- 17 -
5.1.10.
The Hold Function
There are three types of hold functions which can be used for different purposes.
Normal hold
Holds the value displayed at the time the hold command was received.
Peak hold
Holds the maximum value reached after the hold command was received.
The value will be refreshed if it increases again.
Averaging hold
This function averages weighing data over a certain period of time and then holds the result.
It is useful for measuring things that are difficult to weigh such as an animal that won’t settle
down, or for averaging out the weight of an object in an unstable state. In addition, it can
reduce the effects of breezes which the digital filter cannot eliminate.
Functions Related To The Hold Function
(Hold function)
:
The type of hold function can be selected.
To set operating conditions for the hold function
:
Set the averaging time length, standby
time, or start and stop conditions with
to
. (Will not affect normal holds)
Peak hold
Averaging hold
Averaging (Busy state)
(Averaging time)
Hold
Hold start command
Automatic start condition
Averaging start
Averaging finish Hold cancellation
command
Automatic cancellation condition
Standard
state
Standard
state
Hold
Hold start command
Automatic start condition
Hold cancellation command
Automatic cancellation condition
Standard
state
Standard
state
Hold
Judgment start
Illustration 6 Peak hold / Averaging hold