Fault Removal Instructions
8.3.1
Replacement of Inlet Fuse(s)
Fig.
16: Fuses (two) in combi seal at the back of the pipettor
Replacement of system fusing
Fuse replacement is complete.
WARNING
There
is
danger
of
exposure
to
electric
voltage
!
Always
detach
the
power
supply
cable
from
the
line
power
socket
before
you
replace
a
fuse.
Check
for
absense
of
voltage
in
the
pipettor.
Description
Type/Rated specs
Line
power
fuse
T 6,3 A,
250V
AC,
5x20 mm
1.
Turn
power
supply
off.
2.
Detach
line
power
cord
from
system
inlet
port
and
mains
socket.
3.
Use
a
small
screwdriver
to
extract
the
fuse
holder
from
the
combi
seal
4.
Perform
check
to
identify
the
defective
fuse.
5.
Remove
defective
fuse
from
fuse
holder.
6.
Install
replacement
fuse
(T6.3A)
in
fuse
holder.
7.
Restore
the
fuse
holder
to
its
position.
8.
Turn
power
supply
on
again.
Summary of Contents for 845-00003-03
Page 1: ...InnuPure C96 User Manual ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ...2015 VII ...
Page 8: ...VIII 2015 ...
Page 12: ...XII 2015 ...
Page 20: ...8 2015 General Information ...
Page 44: ...32 2015 Technical Description ...
Page 54: ...42 2015 Routine Start Up Procedure ...
Page 60: ...48 2015 Operation ...
Page 64: ...52 2015 What to do if ...
Page 72: ...60 2015 Shutting Down ...
Page 76: ...64 2015 Waste Disposal ...