PortaSens
III
Portable Gas Leak Detector, Model D16
Operation and Maintenance Manual
Rev-A Feb 2018
- 54 -
Configuring Timed Sampling
Table 22 Configuring Timed Sampling.
Description
Display
Sample Time
On the “Timed Sampling” page (see
Table 21
above), touch the “Sample Time” button. The
numeric keypad control will appear and display
the current value, as shown on the right. Use the
keypad to enter the new value, and then touch
the “Ok” button to save it. Touch the “Back”
button to close the control and return without
saving.
Setting range:
10s – 10m
Measure Time
On the “Timed Sampling” page (see
Table 21
above), touch the “Measure Time” button. The
numeric keypad control will appear and display
the current value, as shown on the right. Use the
keypad to enter the new value, and then touch
the “Ok” button to save it. Touch the “Back”
button to close the control and return without
saving.
Setting range:
10s – 10m
Clear Level
On the “Timed Sampling” page (see
Table 21
above), touch the “Clear Level” button. The
numeric keypad control will appear and display
the current value, as shown on the right. Use the
keypad to enter the new value, and then touch
the “Ok” button to save it. Touch the “Back”
button to close the control and return without
saving.
Setting range:
5 – 100 % of sensor “Range”
setting.