PortaSens
III
Portable Gas Leak Detector, Model D16
Operation and Maintenance Manual
Rev-A Feb 2018
- 32 -
Configuring the Pump
Table 11 Configuring the pump.
Description
Display
Pump Control Page
On the “Main Display” page, touch the “Menu”
button, and then touch the “Pump” button. This
will present the “Pump Control” page, as shown
on the right.
The page presents the current pump status:
Normal (or Restricted)
– the pump will operate
at startup, and will maintain 400cc/min (0.85
SCFH).
Disabled
– the pump does not run, even after
the D16 is restarted.
Pump On/Off (Toggle)
On the “Pump Control” page, touch the “On/Off”
button. If the pump status was “Normal”, it will
be changed to “Disabled”. Otherwise, the control
will be changed back to “Normal”.
Pump Duty Cycle
On the “Pump Control” page, touch the “Duty
Cycle %” button and adjust the pump motor
speed up or down to achieve the desired flow
rate.
Setting range
: 1-100 %