PortaSens
III
Portable Gas Leak Detector, Model D16
Operation and Maintenance Manual
Rev-A Feb 2018
- 36 -
Description
Display
Blanking
On the “Sensor Menu” page, touch the “Blanking”
button. The numeric keypad control will appear
and display the current value, as shown on the
right. Use the keypad to enter the new value, and
then touch the “Ok” button to save it. Touch the
“Back” button to close the control and return
without saving.
Setting range
: The blanking value appears in the
sensor’s units of measure (ie, PPM) and may be
set from 0 to 10 % of the “Range” value. Changing
the “Range” value changes the blanking value
proportionally.
Averaging
On the “Sensor Menu” page, touch the “Averaging”
button. The numeric keypad control will appear
and display the current value, as shown on the
right. Use the keypad to enter the new value, and
then touch the “Ok” button to save it. Touch the
“Back” button to close the control and return
without saving.
Setting range
: 1-100 (approximately the time in
seconds to reach T95)