O2NE+ & Safe-Ox+ 
User Manual 

 

Int. Approved   

 

Document ref: P0120-820-14 

August 2017 

Page 5 of 23 

 

 

Commercial in Confidence 

 

Overview 

Two variants of this product are available.  The Ax 

O2NE+™ O2 detector is designed to detect 

the presence of low oxygen in ambient air.  
 
The Ax Safe-O

x+™ O2 detector is designed to detect the presence of high and low oxygen in 

ambient air. 
 

Instrument Name 

Application 

Sensor Type 

Ax 

O2NE+™ 

Fixed instrument configured as standard with two 
depletion  alarms,  representing  warning  (19.5%) 
and danger (18%) levels of oxygen 

Partial pressure 

Ax Safe-

Ox+™ 

Fixed instrument configured as standard with two 
pre-set alarm levels at 23.0% and 19.5% O2 

Partial pressure 

3.1 

Partial pressure sensor

 

This  type  of  sensor  measures  the  partial  pressure  of  oxygen  in  the  atmosphere  where  the 
instrument is located. Therefore it is directly affected by fluctuations in atmospheric pressure. 
To  counteract  this,  a  pressure  compensation  sensor  is  used  to  compensate  for  changes  in 
atmospheric pressure.  
 
The  oxygen  sensor  used  is  not  cross  sensitive  to  helium  so  will  measure  correctly  in  the 
presence of a helium atmosphere.  This means that the sensor can be safely used to detect 
oxygen  displacement  by  helium  gas  leaks.    Care  should  be  taken  however,  if  the  Ax 
O2NE+™/Safe-Ox+™ is to be permanently exposed to a helium environment, as this can affect 
the performance of the pressure compensation sensor over time.  Contact Analox for installation 
advice if you intend to use the unit in this way. 

3.2 

Setup 

Different versions of the instrument allow operation from  

1)

 

210/250V AC supply 

2)

 

110/120V AC supply 

3)

 

9-24V DC supply 

 
The Ax 

O2NE+™/Safe-Ox+™ is intended to be used as a permanent installation. It provides a 

digital readout of oxygen, plus audible and visual alarms to potentially dangerous deficiencies 
(and abundancies in case of Ax Safe-

Ox+™) of oxygen in the air surrounding the instrument. 

The instrument uses an electrochemical cell together with state of the art technology, built in 
an  IP65  splash  proof  housing  and  is  designed  to  provide  long,  trouble  free  service,  with 
minimum maintenance. 
 
The Ax 

O2NE+™ has two pre-set alarm levels at 19.5% and 18% O2. The alarms are set with 

a  small  hysteresis  which  means  the  O2  concentration  has  to  rise  above  the  alarm  set  point 
before the alarm will cancel.  
  
The Ax Safe-

Ox+™ has two pre-set alarm levels at 23.0% and 19.5% O2. The alarms are set 

with a small hysteresis which means the O2 concentration has to fall below the 23.0% alarm 
setpoint or rise above the 19.5% alarm setpoint before the alarm will cancel.  
 
Optional items fitted to or supplied with the unit may include the following: 

1)  Remote alarm repeater* 
2)  One or two medium duty relays 
3)  Test gas, flow indicators and control valves* 
4)  Flow adaptor*   

Items marked with an asterisk (*) do not need to be specified at the time of order and may be 
retro fitted 

Summary of Contents for O2NE+

Page 1: ...usiness Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net O2NE Safe Ox User Manual THIS IS A PRECISION DEVICE KEEP UP RIGHT ALLOW 2 HOUR WARM UP...

Page 2: ......

Page 3: ...splay 6 4 2 Wall mounting 6 4 3 Wiring installation 7 4 4 AC supply models 7 4 5 DC supply models 7 4 6 Alarm repeaters 8 4 7 Alarm relay output models 10 4 8 Relay wiring 10 4 9 4 to 20mA output mode...

Page 4: ...early to 50 relative humidity at 40 C 5 Mains voltage supply fluctuations not to exceed 10 of the nominal voltage 6 Impulse withstand over voltage category II of IEC 60364 4 443 7 Pollution degree 2 8...

Page 5: ...nally connected equipment introducing hazardous voltages into the A50 O2NE Safe Ox or CO Clear It is the responsibility of users of this product to ensure that their equipment accessory is safe before...

Page 6: ...On receipt of the Ax O2NE Safe Ox please check you have the following 1 Ax O2NE Safe Ox main unit and alarm repeater 2 User manual 3 Test certificate 4 Rawl plugs and screws for wall mounting 5 Drill...

Page 7: ...s this can affect the performance of the pressure compensation sensor over time Contact Analox for installation advice if you intend to use the unit in this way 3 2 Setup Different versions of the ins...

Page 8: ...be removed from the instrument display window This is an easy peel label and should be removed easily without leaving behind any residue 4 2 Wall mounting The Ax O2NE Safe Ox should be mounted onto a...

Page 9: ...Safe Ox s are pre wired with a mains cable fitted with a plug suited to the destination country Where internal plug fuses are fitted these are 3 amp Ensure that the unit is connected to the correct su...

Page 10: ...Ax O2NE Safe Ox pass through one of the cable glands on the Remote Alarm Repeater 3 Terminate the wires in accordance with the table below 1 Replace the lid of the Remote Alarm Repeater insert and ti...

Page 11: ...ote Alarm Repeater 3 Terminate the wires in accordance with the table below 4 Connect the free wire Brown or Orange to A50 231 main PCB at JP13 terminal 5 3 Replace the lid of the Remote Alarm Repeate...

Page 12: ...should be connected 4 8 Relay wiring The cable gland is for cables of outside diameter between 5 and 7mm if cable fitted is outside that range a suitably specified cable gland must be used Ensure tha...

Page 13: ...Document ref P0120 820 14 August 2017 Page 11 of 23 Commercial in Confidence The customer system should be connected to this cable as follows and as per the following drawing 1 Blue wire 4 to 20mA ne...

Page 14: ...of alarms They may be factory configured in the opposite sense if required 5 2 Alarm indications 5 2 1 Ax Safe Ox If the Ax Safe Ox detects an O2 concentration which is greater than the high alarm lev...

Page 15: ...wn the instrument and remove from its installed location to allow access to the lid screws Remove the lid of the instrument to gain access to the internals of the instrument Disconnect the sensor from...

Page 16: ...ses used on the Ax O2NE Safe Ox must be manufactured to a 2 tolerance Note For best performance the monitor should be powered and allowed to warm up for approximately 2 hours before calibration is per...

Page 17: ...will illuminate sound 4 times before returning to normal operation If during the one minute wait period the red or amber LED s are illuminated consult the table below ALARM1 LAMP RED ALARM2 LAMP RED F...

Page 18: ...w for the flow of gas over the sensor 5 After a few moments the Alarm 1 alarm should operate 6 Remove the test gas It will take a little while for the gas to diffuse out of the instrument sensor allow...

Page 19: ...ND FAST BUZZER O2 Cell Fault Output too High OFF FLASHING FLASHING FLASHING AND FAST BUZZER O2 Cell Fault Output too Low OFF OFF OFF FLASHING AND VERY FAST BUZZER Internal system fault contact Analox...

Page 20: ...O2NE 600g Alarm Repeater 150g Dimensions O2NE Safe Ox 175x105x75 mm Alarm Repeater 155x72x45 mm IP rating O2NE Safe Ox IP65 Alarm Repeater Hard wired IP65 Quick connect IP43 Sensor type Electro Chemi...

Page 21: ...14 August 2017 Page 19 of 23 Commercial in Confidence 8 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for info...

Page 22: ...O2NE Safe Ox User Manual Int Approved Document ref P0120 820 14 August 2017 Page 20 of 23 Commercial in Confidence 9 Declaration of conformity...