Safety information  

Every gas monitor installation should be risk assessed.  The correct location of monitor(s) must 
consider the potential sources of gas leaks and the location of expected human occupation in 
the  area.      Where  large  areas  must  be  monitored,  it  is  often  advised  that  no  single  monitor 
should  cover  a  volume  in  excess  of  80m

3

.    The  installation  of  more  monitors  should  be 

considered where an area is complex in shape, filled with obstacles, has a lack of ventilation or 
air circulation, or if there are dead zones where gas can collect 
 
The  Ax 

O2NE+™/Safe-Ox+™  is  designed  to  be  compliant  with  the  following  standards: 

EN61010-1:  2001,  IEC61010-1:  2001.  It  is  designed  to  be  safe  at  least  under  the  following 
conditions. 
 

1)  Indoor use 

 

2)  Altitude up to 1800m 

 

3)  Temperature 0°C to +40°C 

 

4)  Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% 

relative humidity at 40°C. 
 

5)  Mains voltage supply fluctuations not to exceed 10% of the nominal voltage 

 

6)  Impulse withstand (over-voltage) category II of IEC 60364-4-443 

 

7)  Pollution degree 2 

 

8)  Mains voltage:- 

a.  230V AC  

(Not adjustable - Instrument will be factory set) 

b.  110V AC  

(Not adjustable - Instrument will be factory set) 

c.  24V DC  

(Not adjustable - Instrument will be factory set) 

 

9)  Mains power:- 

a.  Less than 5VA 

– 110V AC and 230V AC versions 

b.  Less than 5W 

– 24V DC version. 

 

10) Mains frequency - 50/60Hz 

 

11) The remote alarm repeater has ingress protection to IP43: direct sprays of water up to 

60°  from  the  vertical  in  accordance  with  EN  60529:1991  +  A1.    All  other  units  have 
ingress  protection  to  IP65:  low  pressure  water  jets  from  all  directions  and  totally 
protected from dust in accordance with EN 60529:1991 + A1. 
 

12) Insulation: - Reinforced insulation, class II product according to IEC536. 

 

13) Not for use in corrosive or explosive atmospheres 

 

14) Not approved for use in vehicles, ships or aircraft 

 
Fuse ratings:- 
230V AC, 500mA,  

F rating 250V (20mm x 5mm glass cartridge) 

110V AC, 500mA,  

F rating 250V (20mm x 5mm glass cartridge) 

9-24V DC, 200mA,  

AS rating 250V (20mm x 5mm glass cartridge) 

 
 
 
 

Summary of Contents for O2NE+

Page 1: ...usiness Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net O2NE Safe Ox User Manual THIS IS A PRECISION DEVICE KEEP UP RIGHT ALLOW 2 HOUR WARM UP...

Page 2: ......

Page 3: ...splay 6 4 2 Wall mounting 6 4 3 Wiring installation 7 4 4 AC supply models 7 4 5 DC supply models 7 4 6 Alarm repeaters 8 4 7 Alarm relay output models 10 4 8 Relay wiring 10 4 9 4 to 20mA output mode...

Page 4: ...early to 50 relative humidity at 40 C 5 Mains voltage supply fluctuations not to exceed 10 of the nominal voltage 6 Impulse withstand over voltage category II of IEC 60364 4 443 7 Pollution degree 2 8...

Page 5: ...nally connected equipment introducing hazardous voltages into the A50 O2NE Safe Ox or CO Clear It is the responsibility of users of this product to ensure that their equipment accessory is safe before...

Page 6: ...On receipt of the Ax O2NE Safe Ox please check you have the following 1 Ax O2NE Safe Ox main unit and alarm repeater 2 User manual 3 Test certificate 4 Rawl plugs and screws for wall mounting 5 Drill...

Page 7: ...s this can affect the performance of the pressure compensation sensor over time Contact Analox for installation advice if you intend to use the unit in this way 3 2 Setup Different versions of the ins...

Page 8: ...be removed from the instrument display window This is an easy peel label and should be removed easily without leaving behind any residue 4 2 Wall mounting The Ax O2NE Safe Ox should be mounted onto a...

Page 9: ...Safe Ox s are pre wired with a mains cable fitted with a plug suited to the destination country Where internal plug fuses are fitted these are 3 amp Ensure that the unit is connected to the correct su...

Page 10: ...Ax O2NE Safe Ox pass through one of the cable glands on the Remote Alarm Repeater 3 Terminate the wires in accordance with the table below 1 Replace the lid of the Remote Alarm Repeater insert and ti...

Page 11: ...ote Alarm Repeater 3 Terminate the wires in accordance with the table below 4 Connect the free wire Brown or Orange to A50 231 main PCB at JP13 terminal 5 3 Replace the lid of the Remote Alarm Repeate...

Page 12: ...should be connected 4 8 Relay wiring The cable gland is for cables of outside diameter between 5 and 7mm if cable fitted is outside that range a suitably specified cable gland must be used Ensure tha...

Page 13: ...Document ref P0120 820 14 August 2017 Page 11 of 23 Commercial in Confidence The customer system should be connected to this cable as follows and as per the following drawing 1 Blue wire 4 to 20mA ne...

Page 14: ...of alarms They may be factory configured in the opposite sense if required 5 2 Alarm indications 5 2 1 Ax Safe Ox If the Ax Safe Ox detects an O2 concentration which is greater than the high alarm lev...

Page 15: ...wn the instrument and remove from its installed location to allow access to the lid screws Remove the lid of the instrument to gain access to the internals of the instrument Disconnect the sensor from...

Page 16: ...ses used on the Ax O2NE Safe Ox must be manufactured to a 2 tolerance Note For best performance the monitor should be powered and allowed to warm up for approximately 2 hours before calibration is per...

Page 17: ...will illuminate sound 4 times before returning to normal operation If during the one minute wait period the red or amber LED s are illuminated consult the table below ALARM1 LAMP RED ALARM2 LAMP RED F...

Page 18: ...w for the flow of gas over the sensor 5 After a few moments the Alarm 1 alarm should operate 6 Remove the test gas It will take a little while for the gas to diffuse out of the instrument sensor allow...

Page 19: ...ND FAST BUZZER O2 Cell Fault Output too High OFF FLASHING FLASHING FLASHING AND FAST BUZZER O2 Cell Fault Output too Low OFF OFF OFF FLASHING AND VERY FAST BUZZER Internal system fault contact Analox...

Page 20: ...O2NE 600g Alarm Repeater 150g Dimensions O2NE Safe Ox 175x105x75 mm Alarm Repeater 155x72x45 mm IP rating O2NE Safe Ox IP65 Alarm Repeater Hard wired IP65 Quick connect IP43 Sensor type Electro Chemi...

Page 21: ...14 August 2017 Page 19 of 23 Commercial in Confidence 8 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for info...

Page 22: ...O2NE Safe Ox User Manual Int Approved Document ref P0120 820 14 August 2017 Page 20 of 23 Commercial in Confidence 9 Declaration of conformity...

Page 23: ......

Reviews: