4.6
Alarm repeaters
The Alarm repeater has four status indicators and a Mode button, which mimic the button and
indicators on the main Ax
O2NE+™/Safe-Ox+™ enclosure.
An 8 metre, 8 core cable is pre-wired to the Ax
O2NE+™/Safe-Ox+™ on units to be fitted with
an alarm repeater. This ensures that for a basic installation, there is no need to dismantle the
Ax
O2NE+™/Safe-Ox+™ main unit. A maximum of three repeaters may be fitted in a daisy
chain configuration.
4.6.1
To connect a “Hard Wired” Remote Alarm Repeater
1) Disconnect the power supply from the Ax
O2NE+™/Safe-Ox+™.
2) Open the Remote Alarm Repeater unit by removing the 4 screws in the front of the
case and carefully pulling the case apart. The connecting wires from the Ax
O2NE+™/Safe-Ox+™ pass through one of the cable glands on the Remote Alarm
Repeater.
3) Terminate the wires in accordance with the table below
1) Replace the lid of the Remote Alarm Repeater, insert and tighten the 4 screws, and
mount it in the desired position using the 2 fixing lugs.
2) Restore power to the Ax
O2NE+™/Safe-Ox+™. Press the switch on the Remote
Alarm Repeater once, and ensure that the four indicators flash. Note that in the
presence of a genuine alarm, this test feature is disabled.
4.6.2
To connect a “Quick Connect” Remote Alarm Repeater
1) Disconnect the power supply from the Ax
O2NE+™/Safe-Ox+™.
2) Insert the connector on the end of the wire into either socket on the bottom of the
Remote Alarm Repeater.
3) Restore power to the Ax
O2NE+™/Safe-Ox+™. Press the switch on the Remote Alarm
Repeater once, and ensure that the four indicators flash. Note that in the presence of
a genuine alarm, this test feature is disabled.
REPEATER CABLE COLOURS
Type 1
Repeater
Term
Type 2 (Cat 5)
ST1
Terminal
No
8
3
7
2
6
4
5
5
4
6
3
7
2
8
1
1