O2NE+ & Safe-Ox+ 
User Manual 

 

Int. Approved   

 

Document ref: P0120-820-14 

August 2017 

Page 15 of 23 

 

 

Commercial in Confidence 

 

2)  Enter Technician Mode by pressing the mode switch 3 times.  If entered successfully 

the green LED will flash off for 1.5 seconds and on for 0.5 of a second. 

 
3) 

Select auto calibration by pressing the mode switch 4 times.  The 2 Red Alarm LED’s 
will light up to show you are now in this mode. 

 

4) 

Press the mode switch 2 times to start the Auto Calibration, the Red Alarm LED’s will 
turn off and the Green LED will continue to flash. 

 

5)  Wait one minute for the instrument to adjust. When the instrument has a new 

calibration value, the buzzer will sound one bleep and all the LED’s will be off. 

 

6)  Accept this new calibration value by pressing the mode switch 2 times. The green 

LED will flash to show the instrument has accepted the new oxygen value. 

 
To return to normal operation, press the mode switch once. The LED’s and buzzer will illuminate 
/ sound 4 times before returning to normal operation. 
 
If, during the 

one minute wait period, the red or amber LED’s are illuminated consult the table 

below. 

 

ALARM1 LAMP 

(RED) 

 

ALARM2 LAMP 

(RED) 

 

FAULT LAMP 

(YELLOW) 

 

MEANING 

 

 

ACTION 

 

 

ON 

 

OFF 

 

OFF 

 

Calibration gas is 

above allowable 

range 

Ensure the 

calibration gas is 
20.9% oxygen in 

nitrogen with      

+/-2% tolerance 

 

OFF 

 

ON 

 

OFF 

 

Calibration gas is 

below allowable 

range 

Ensure the 

calibration gas is 
20.9% oxygen in 

nitrogen with      

+/-2% tolerance 

 

OFF 

 

OFF 

 

ON 

 

Internal fault 

Contact Analox. 

 
Note:  If  at any  time  you  send  the  wrong  instruction  and  would  like  to abort,  press  the  mode 
switch  once  and  wait  20  seconds,  this  should  bring  you  back  to technician  mode. Press  the 
mode switch once again, this should bring you back to normal operation, th

e LED’s and buzzer 

will  illuminate/sound  4  times.    Alternatively,  disconnect  the  power  supply  to  the  Ax 
O2NE+™/Safe-Ox+™, wait a moment and re-connect power. 
 

6.3 

Alarm checks 

To verify that the indicators and the audible alarms are working, press the Mode switch on the 
Ax 

O2NE+™/Safe-Ox+™  or  any  of  its  repeaters.    The  indicators  and  the  audible  alarm  will 

pulse four times. 
 

 

Summary of Contents for O2NE+

Page 1: ...usiness Park North Yorkshire TS9 5PT UK T 44 0 1642 711400 F 44 0 1642 713900 W www analox net E info analox net O2NE Safe Ox User Manual THIS IS A PRECISION DEVICE KEEP UP RIGHT ALLOW 2 HOUR WARM UP...

Page 2: ......

Page 3: ...splay 6 4 2 Wall mounting 6 4 3 Wiring installation 7 4 4 AC supply models 7 4 5 DC supply models 7 4 6 Alarm repeaters 8 4 7 Alarm relay output models 10 4 8 Relay wiring 10 4 9 4 to 20mA output mode...

Page 4: ...early to 50 relative humidity at 40 C 5 Mains voltage supply fluctuations not to exceed 10 of the nominal voltage 6 Impulse withstand over voltage category II of IEC 60364 4 443 7 Pollution degree 2 8...

Page 5: ...nally connected equipment introducing hazardous voltages into the A50 O2NE Safe Ox or CO Clear It is the responsibility of users of this product to ensure that their equipment accessory is safe before...

Page 6: ...On receipt of the Ax O2NE Safe Ox please check you have the following 1 Ax O2NE Safe Ox main unit and alarm repeater 2 User manual 3 Test certificate 4 Rawl plugs and screws for wall mounting 5 Drill...

Page 7: ...s this can affect the performance of the pressure compensation sensor over time Contact Analox for installation advice if you intend to use the unit in this way 3 2 Setup Different versions of the ins...

Page 8: ...be removed from the instrument display window This is an easy peel label and should be removed easily without leaving behind any residue 4 2 Wall mounting The Ax O2NE Safe Ox should be mounted onto a...

Page 9: ...Safe Ox s are pre wired with a mains cable fitted with a plug suited to the destination country Where internal plug fuses are fitted these are 3 amp Ensure that the unit is connected to the correct su...

Page 10: ...Ax O2NE Safe Ox pass through one of the cable glands on the Remote Alarm Repeater 3 Terminate the wires in accordance with the table below 1 Replace the lid of the Remote Alarm Repeater insert and ti...

Page 11: ...ote Alarm Repeater 3 Terminate the wires in accordance with the table below 4 Connect the free wire Brown or Orange to A50 231 main PCB at JP13 terminal 5 3 Replace the lid of the Remote Alarm Repeate...

Page 12: ...should be connected 4 8 Relay wiring The cable gland is for cables of outside diameter between 5 and 7mm if cable fitted is outside that range a suitably specified cable gland must be used Ensure tha...

Page 13: ...Document ref P0120 820 14 August 2017 Page 11 of 23 Commercial in Confidence The customer system should be connected to this cable as follows and as per the following drawing 1 Blue wire 4 to 20mA ne...

Page 14: ...of alarms They may be factory configured in the opposite sense if required 5 2 Alarm indications 5 2 1 Ax Safe Ox If the Ax Safe Ox detects an O2 concentration which is greater than the high alarm lev...

Page 15: ...wn the instrument and remove from its installed location to allow access to the lid screws Remove the lid of the instrument to gain access to the internals of the instrument Disconnect the sensor from...

Page 16: ...ses used on the Ax O2NE Safe Ox must be manufactured to a 2 tolerance Note For best performance the monitor should be powered and allowed to warm up for approximately 2 hours before calibration is per...

Page 17: ...will illuminate sound 4 times before returning to normal operation If during the one minute wait period the red or amber LED s are illuminated consult the table below ALARM1 LAMP RED ALARM2 LAMP RED F...

Page 18: ...w for the flow of gas over the sensor 5 After a few moments the Alarm 1 alarm should operate 6 Remove the test gas It will take a little while for the gas to diffuse out of the instrument sensor allow...

Page 19: ...ND FAST BUZZER O2 Cell Fault Output too High OFF FLASHING FLASHING FLASHING AND FAST BUZZER O2 Cell Fault Output too Low OFF OFF OFF FLASHING AND VERY FAST BUZZER Internal system fault contact Analox...

Page 20: ...O2NE 600g Alarm Repeater 150g Dimensions O2NE Safe Ox 175x105x75 mm Alarm Repeater 155x72x45 mm IP rating O2NE Safe Ox IP65 Alarm Repeater Hard wired IP65 Quick connect IP43 Sensor type Electro Chemi...

Page 21: ...14 August 2017 Page 19 of 23 Commercial in Confidence 8 Disposal According to WEEE regulation this electronic product can not be placed in household waste bins Please check local regulations for info...

Page 22: ...O2NE Safe Ox User Manual Int Approved Document ref P0120 820 14 August 2017 Page 20 of 23 Commercial in Confidence 9 Declaration of conformity...

Page 23: ......

Reviews: