ANALOX O2 PORTABLE OXYGEN ANALYSER
Operators Manual MO2-807-02 January 2009
Page 10
7.0
Maintenance
7.1
Battery replacement.
a)
Remove the 4 screws located at each corner of the unit and carefully
lift the lid.
b)
Slide the battery out of its spring bracket and disconnect the lead.
c)
Connect the lead to the new battery and slide the battery behind the
spring bracket.
d)
Replace the lid carefully and screw down taking care that the sensor
locates properly.
e)
Ensure that you do not trap any wires.
7.2
Sensor replacement.
a)
Replacement sensor part number: 9100-9212-5AD
b)
Remove the 4 screws located at each corner of the unit and carefully
lift the lid.
c)
Remove the flow adapter if fitted and slide the sensor out of the lid.
d)
Disconnect the in line sensor connector.
e)
Dispose of the old sensor according to local regulations for Lead and
Potassium Hydroxide solution.
f)
Remove the new sensor from its bag and check it for leaks, connect
to the inline connector and slide through the lid.
g)
Replace the lid carefully and screw down taking care that the sensor
locates properly. Ensure that you do not trap any wires.