4.0
Pure Oxygen calibration for measuring oxygen purity up to
100%
.
4.1 Connect 100% bottled oxygen ( certified ) to the flow adaptor
and adjust the flowrate to between 0.5 and 1.0 litre per minute.
4.2
Allow the reading to settle.
4.3
Expose to atmosphere, the reading should display +/- 0.4 of
corrected humidity value i.e. between 20.5 and 21.3