UG-1882
Rev. 0 | Page 19 of 28
25058-
023
Figure 21.
Histogram
Tab
25058-
024
Figure 22.
FFT
Tab
Capture Pane
General Capture Settings
The
Sequence Count
list box allows the user to select the
number of times to execute the configured channel sequence
per capture.
The user can enter the sample rate in kilosamples per second
(kSPS) in the
Throughput (kSPS)
box. Note that this box sets
the sample rate of the
ADC core (for example, CNV
signal frequency), whereas the effective sample rate for each
AD4696 channel depends on the sequencer configuration.
Device Configuration
The
Sequencer Mode
dropdown box allows the user to select
the desired sequencer mode.
Clicking the
Go to Sequencer
button opens either the standard
sequencer view or the advanced sequencer view based on the
value selected in the
Sequencer Mode
dropdown menu.
Reference Settings
Use the
Reference Voltage
text box to enter the value of the
reference voltage on the connected hardware. The
Reference
Voltage
box is used by the analysis math to convert between
ADC codes and equivalent input voltage. Changing the
Reference
Voltage
value does not modify any hardware or memory map
settings on the connected hardware. If the value in the
Reference
Voltage
box does not match the reference voltage used on the
connected hardware, the data in the
Waveform
graph is
incorrect when displayed in volts.