720A Operation and Maintenance Manual – Rev S – January 25, 2019
Page 27
13231 Rooster Springs Rd.,
Austin, TX 78737
T:
(512) 858-4045
E:
amptec.com
Trim pot at RV1:
Influences full-scale on all ranges.
Trim pot at RV2:
Influences zero-scale on all ranges.
Trim pot at RV5:
Influences zero-scale on 200 m
Ω
range.
Trim pot at RV6:
Influences full-scale on 200 m
Ω
range.
All other trim pots on the board, including RV3 and RV201 shown in
Figure 9
, are not
used during standard product recalibration and should not be adjusted during this
procedure.
E-4.3 Calibration Procedure Adjustments
This section contains information on the specific adjustments necessary to calibrate a
720A meter. Ensure that you have first followed the instructions detailed in the
calibration procedure overview in section X before completing any of the internal
adjustments below. It is recommended that you complete the following steps in order.
1. Zero Adjustment:
Select the 200
Ω
range. Connect the meter via the
measurement test leads across a precision 1.0
Ω
resistance standard with ± 0.005% or better accuracy.
Adjust the trim-pot (RV2) to make the meter read 1.00
Ω
(100 counts on the display). Note that this process
will be repeated on the 200
mΩ
range with a 0.001
Ω
standard
after
the full-scale adjustment is performed.
2. Full-Scale Adjustment:
Connect the meter via the measurement test leads
across a precision 100
Ω
resistance standard with ±
0.005% or better accuracy. Adjust the full-scale pot
(RV1) to make the meter read 100.00
Ω
(10000
counts on the display).
3. 200 m
Ω
Adjustment:
Next, the 200 m
Ω
range needs to be calibrated. The
trim pots adjusted above (RV1 & RV2) affect every
range of the meter, while RV5 & RV6 only affect the
200 m
Ω
range. Therefore, they must be adjusted last.
4. 200 m
Ω
Zero Adjustment:
Select the 200 m
Ω
range. Connect the meter
via the measurement test leads across a precision
0.001
Ω
resistance standard with ± 0.01% or better
accuracy. Adjust the trim-pot (RV5) to make the meter
read 01.00 mΩ (100 counts on the display).
5. 200 m
Ω
Full-Scale Adjustment:
Connect the meter via the measurement test leads
across a precision 100 m
Ω
resistance standard with ±
0.005% or better accuracy. Adjust the full-scale pot
(RV6) to make the meter read 100.00 m
Ω
(10000
counts on the display).