EMC GUIDANCE & MANUFACTURER’S DECLARATION
32
D-0115690-C (OM001)
– Otowave 102 Operating Manual
18.
EMC GUIDANCE & MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration – electromagnetic emissions
The Otowave 102 Tympanometer is intended for use in the electromagnetic environment specified below.
The customer or user of the Otowave 102 Tympanometer should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment
– guidance
RF emissions
CISPR 11
Group 1
The Otowave 102 Tympanometer uses RF
energy only for its internal function. Therefore,
its RF emissions are very low and are not likely
to cause interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The Otowave 102 Tympanometer is suitable for
use in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration – electromagnetic immunity (1)
The Otowave 102 Tympanometer is intended for use in the electromagnetic environment specified below.
The customer or user of the Otowave 102 Tympanometer should assure that it is used in such an
environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic
environment
– guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors
should
be
wood,
concrete or ceramic tile. If
floors
are
covered
with
synthetic material, the relative
humidity should be at least
30%
Electrical
fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/output
lines
Not applicable
Not applicable
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common mode
Not applicable
Not applicable