ERROR MESSAGES & FAULT CONDITIONS
25
D-0115690-C (OM001)
– Otowave 102 Operating Manual
14.
ERROR MESSAGES & FAULT CONDITIONS
If a fault condition cannot be cleared, the operator is cautioned against repeatedly starting the
instrument.
Please note:
Refer to the installation & operating instructions provided with the NOAH
Impedance Module or ampliSuite software for details of the data transfer operation and errors that
may occur.
PROBLEM
CAUSE
SOLUTION(S)
My profiles names are gone. The change of the operation
language
will
automatically
default the profile names to the
factory
description.
It
is
important to note that the
settings of each profile stay the
same.
Re-name the profile names
again.
No pressure can be build up
and the test sequence will
stay in the
EQUALIZE
PRESSURE SCREEN
.
No seal can be obtained
Estimated volume is too high
(perforated ear drum)
Wrong ear tip size chosen
Probe is blocked
Examine the probe tip for
contamination and clean or
replace the probe tip
Reposition the probe
Change the ear tip
No reflex test is conducted
after the tympanometry even
though the reflex test is
active
in
the
REFLEX
SEQUENCE.
In
REFLEX SELECTION
the
setting is set to
ONLY IF PEAK
IS
FOUND
or
NEVER
MEASURE.
Change settings in
REFLEX
SELECTION
to desired option.
Last measured data cannot
be found under
VIEW THE
LAST TEST
.
NEW TEST
might have been
selected
in
between
and
thereby deleted the last test(s)
from the short-term memory.
To be stored data should be
stored immediately.
BLOCKED PROBE
Indicator LED b and c flash
fast.
Probe is blocked
Probe placed against ear
canal skin
Probe
disconnected
from
base unit
Examine the probe tip for
contamination and clean or
replace the probe tip
Reposition the probe
Change the ear tip
Check probe connection with
base unit
WITHDRAW PROBE
Indicator LED b and c flash
fast.
The probe has been moved
during measurement.
Test has been started with the
probe already inserted into
the ear.
Reposition the probe
Volume outside range
WITHDRAW PROBE
Indicator LED b and c flash
fast.
Ear canal volume is > 5ml.
Probe is not properly inserted
into the ear.
Reposition the probe