CONFIGURATION
14
D-0115690-C (OM001)
– Otowave 102 Operating Manual
6.
CONFIGURATION
6.1.
SWEEP SETTINGS
Video available on how to add ipsilateral reflexes to test protocol.
ITEM
DESCRIPTION
DEFAULT
Test Sequence:
When testing both ears, define what ear side the test will start
with.
R, L
Ear Seal:
The
STANDARD
option is adequate for most tests, although it
may not always be possible to generate the extremes of
pressure during a tympanogram measurement with this
setting.
If difficulty is experienced in using the ear tips to create a seal
the alternative
EXTENDED
option may be helpful. This
function checks that a range of pressures will be available
before starting a test by means of a visual indication of the
quality of the seal.
The
EXTENDED
function is especially helpful if small ear canal
volumes should not experience excessive pressure.
Standard
Defaults:
Reset the sweep settings of the selected profile(s) to its original
settings.
6.2.
REFLEX OPTIONS
Video available on how to add ipsilateral reflexes to test protocol.
ITEM
DESCRIPTION
DEFAULT
Level Mode:
Please note:
Depending on the
LEVEL MODE
selection,
the
LEVELS
screen will contain different content.
ONE LEVEL:
Use the
S keys
to choose the level of reflex
stimulus to apply. Only one level will be tested in the
measurement. The maximum level of ipsilateral stimulus may
be set to maximum 100dBHL; the maximum level of
contralateral stimulus may be set to 110dBHL.
MULTILEVEL:
Use the
S keys
to choose the maximum level
of reflex stimulus to apply and the step size between the levels
of the preceding stimuli. The maximum level of ipsilateral
Multilevel