8.
Remove the nameplate by removing the 4 x panhead screws with a 5/64” hex wrench.
Name Plate
9.
Disconnect the detector cable from the detector PCB. Tuck away the cable so it is out of the way.
10.
Remove the (4) socket head cap screws of the Herriot cell near end cap with a 7/64 hex wrench.
Utilizing the pull bracket, pull the end cap directly away from the Herriot cell block. Set the endcap
down with the mirror facing up, protecting the mirror with supplied lens wipe. Discard the endcap
O- ring and prepare to clean the mirrors.
Bracket
Socket Head Cap Screw
Detector Cable
Be careful not to damage dowel pins of Herriot cell when
removing end caps. Proper alignment of the laser during re-
assembly is dependent on the dowel pins.
WARNING
42