![AMI The Barracuda 4010LX Operator'S Manual Download Page 28](http://html.mh-extra.com/html/ami/the-barracuda-4010lx/the-barracuda-4010lx_operators-manual_2932178028.webp)
a) Set the ALARM SETPOINTS.
Enter your desired value for each setpoint and then press the
ENTER key on your laptop. Keep in mind that your values
cannot exceed the limit of the selected analog Output Range
that you previously selected.
Both Alarms have a 1% hysteresis band that correlates with the
measurement range.
Step 4: Alarm Logic & Setup
The Analyzer features 2 independent Moisture Concentration Alarms –
one for ALARM 1 and one for ALARM 2. The settings for these alarms,
including setpoints, relay contacts, close/open logic and alarm delays,
are adjusted through the
COMMAND CENTER
.
It is important that you plan out how you want your ALARM LOGIC to
work for each ALARM before you start adjusting the settings discussed in
this section.
b) Set the ALARM DELAYS.
There are 2 ALARM DELAYS. Each ALARM DELAY setting is
located beneath the corresponding ALARM that it controls.
Enter your desired time duration for each ALARM DELAY and
press the ENTER key on your laptop. You can also adjust using
the UP and DOWN ARROWS. The range is from 0 to 300
minutes.
*This feature is especially helpful at custody transfer points when
customers are allowed to exceed contractual limits for a
predetermined amount of time.
27