42
8.5 EUT fail input
Between pin 6 (hi) and pin 2, 8, 15, 20 (low)
Input open = inactive; input shorted = active
This connection serves as a control input that can be activated externally.
The EUT can activate this input if it is capable of reporting a disturbance effect caused during an EMC
test. Such events are time / date stamped by the system and are stored together with the current test
parameters for subsequent use in a test report if required.
8.6 EUT power off
Between pin 4 (hi) and pin 2, 8, 15, 20 (low)
Input open =
Inactive, EUT power is controlled via front panel or WIN 3000 software
Input shorted = Active, in case EUT power is switched on, shorting this input will set EUT power to off
Notes:
1. Using this function only makes sense if an EUT power contactor is available somewhere in the
system. EUT power contactors are available in CDN 3153-S63.
2. First the EUT power has to be switched ON via front panel or WIN 3000 software. This way
allows dual drive, as the EUT power can then be switched OFF either from software control or
from this external drive.
3. This signal is also used to drive the orange lamp of INA 3001 warning lamps.
8.7 High voltage active
Between pin 16 (Hi) and pin 2, 8, 15, 20 (low).
This function is activated for firmware revisions 2.30 and higher.
This output is to drive external warning lamps INA 3001. The HV on signal is working together with
the high voltage LED located on the front panel.
Output high (24 V): High voltage is ON
Output low (0 V): High voltage is OFF